Comparison of Mass Discrimination Correction Methods and Sample Introduction Systems for the Determination of Lead Isotopic Composition Using a Multicollector Inductively Coupled Plasma Mass Spectrometer
Authors:
- 21 October 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 80 (22) , 8355-8363
Abstract
The influence of sample introduction system on Neptune MC-ICPMS lead isotopic ratio measurements was tested on dilute solutions of the lead certified material NIST SRM 981 ([Pb] = 0.2−170 ng g−1) using (1) a SIS spray chamber, (2) a MCN 6000 desolvating system, or (3) an Apex inlet system. The impact of using a high-efficiency X-cone in place of a standard H-cone with the MCN and Apex was also investigated. Performance of the sample introduction systems varied with lead concentration. Over 10 ng g−1, no system was significantly more precise or accurate. As lead concentrations decreased, both accuracy and precision diminished, and below 1 ng g−1, use of an X-cone in combination with the Apex and particularly with the MCN system notably improved the quality of the measurements. Various mathematical methods of mass bias correction using thallium additions were tested. Selection of 205Tl/203Tl for NIST SRM 997 to optimize data (1) daily, (2) for each introduction system, and (3) over all sessions significantly improved the data, with no major difference in the output between the three methods. Consistency of the 205Tl/203Tl ratio (2.3888) optimized over all data with previous observations by others supports the use of this value for future measurements.Keywords
This publication has 20 references indexed in Scilit:
- Pb isotopic analysis of standards and samples using a 207Pb–204Pb double spike and thallium to correct for mass bias with a double-focusing MC-ICP-MSChemical Geology, 2004
- Direct Determination of Lead Isotopes (206Pb, 207Pb, 208Pb) in Arctic Ice Samples at Picogram per Gram Levels Using Inductively Coupled Plasma-Sector Field MS Coupled with a High-Efficiency Sample Introduction SystemAnalytical Chemistry, 2004
- A simple method for obtaining highly accurate Pb isotope data by MC-ICP-MSJournal of Analytical Atomic Spectrometry, 2002
- Applications of accurate, high-precision Pb isotope ratio measurement by multi-collector ICP-MSChemical Geology, 2002
- Investigation of matrix effects for Pb isotope ratio measurements by multiple collector ICP-MS: verification and application of optimized analytical protocolsJournal of Analytical Atomic Spectrometry, 2000
- High-precision analysis of Pb isotope ratios by multi-collector ICP-MSChemical Geology, 2000
- Inter-laboratory and other errors in Pb isotope analyses investigated using a 207Pb–204Pb double spikeChemical Geology, 2000
- Precise analysis of copper and zinc isotopic compositions by plasma-source mass spectrometryChemical Geology, 1999
- Lead isotopic analyses of NIST Standard Reference Materials using multiple collector inductively coupled plasma mass spectrometry coupled with a modified external correction method for mass discrimination effectThe Analyst, 1996
- Communication. Aerosol-phase assisted sample digestion for the determination of trace metals in organic samples by plasma atomic emission spectrometryJournal of Analytical Atomic Spectrometry, 1996