Validation of RP-HPLC Method and Stress Degradation for the Combination of Metformin HCl, Atorvastatin Calcium and Glimepiride: Application to Nanoparticles
Open Access
- 12 June 2015
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Chromatographic Science
- Vol. 53 (10) , 1654-1662
Abstract
A stability-indicating high-performance liquid chromatography (HPLC) procedure was developed for the determination of metformin HCl (MTH), atorvastatin calcium (AC) and glimepiride (GP) in combination and their main degradation products. The separation and quantization were achieved on a 5-µm Qualisil gold, C18 column (4.6 mm × 250 mm). The mobile phase selected was phosphate buffer (pH 2.9)–organic phase in proportion of 70:30. Organic phase consisted of methanol–acetonitrile (90:10) at a flow rate of 1 mL/min and detection of analytes was carried out at 230 nm. The method exhibited good linearity over the range of 10–60 µg/mL for MTH, 2–20 µg/mL for AC and 5–30 µg/mL for GP. Square of the correlation coefficients was found to be >0.999. Various stress degradation studies were carried out in combination as per International Conference of Harmonization (ICH) guidelines for 4 h. The recovery and precision were determined in terms of intraday and interday precisions and expressed as relative standard deviations. These were <1 and <2%, respectively. Finally, the applicability of the method was evaluated in nanoparticle analysis of MTH, AC and GP as well as in stability studies of nanoformulation.Keywords
Funding Information
- University Grants Commission AICTE
- University Grants Commission AICTE
This publication has 12 references indexed in Scilit:
- Comparison of ultraviolet detection, evaporative light scattering detection and charged aerosol detection methods for liquid-chromatographic determination of anti-diabetic drugsJournal of Pharmaceutical and Biomedical Analysis, 2010
- Simple and rapid method determination for metformin in human plasma using high performance liquid chromatography tandem mass spectrometry: Application to pharmacokinetic studiesJournal of Chromatography B, 2007
- Development of a RP-HPLC method for screening potentially counterfeit anti-diabetic drugsJournal of Chromatography B, 2007
- Determination of metformin in human plasma by high-performance liquid chromatographyJournal of Chromatography B, 2005
- HPLC study of glimepiride under hydrolytic stress conditionsJournal of Pharmaceutical and Biomedical Analysis, 2004
- Robustness of models developed by multivariate calibration. Part ITrAC Trends in Analytical Chemistry, 2004
- Determination of metformin in plasma using a new ion pair solid phase extraction technique and ion pair liquid chromatographyJournal of Chromatography B, 2003
- Determination of metformin in human plasma by high-performance liquid chromatography with spectrophotometric detectionJournal of Chromatography B: Biomedical Sciences and Applications, 2001
- HPLC-DAD determination of Metformin in human plasma using derivatization with p-nitrobenzoyl chloride in a biphasic systemMicrochemical Journal, 2001
- Ruggedness testing in analytical chemistryTrAC Trends in Analytical Chemistry, 1988