An assessment of the performance of the Baxter elastomeric (LV10) Infusor™ pump under hyperbaric conditions
Authors:
- 31 March 2017
- journal article
- research article
- Published by Diving and Hyperbaric Medicine Journal in Diving and Hyperbaric Medicine Journal
- Vol. 47 (1) , 33-37
Abstract
Introduction: There are limited data on the use of elastomeric infusion pumps during hyperbaric oxygen treatment. Aim: This study evaluated the flow rate of the Baxter elastomeric LV10 InfusorTM pump under normobaric (101.3 kPa and three hyperbaric conditions of 203 kPa, 243 kPa and 284 kPa. Methods: Elastomeric pumps were secured to participants in the same manner as for a typical patient, except that a container collected the delivered antibiotic solution. Pumps and tubing were weighed before and after the test period to determine volume delivered and to calculate flow rates at sea level and the three commonly used hyperbaric treatment pressures at two different time periods, 0–2 hours (h) and 19–21 h into the infusion. Results: The mean flow rates in ml∙hr-1 (SD) were: 9.5 (0.4), 10.3 (0.6), 10.4 (0.6), 10.4 (0.5) at 0−2 h and 10.5 (1.0), 12.2 (0.6), 9.4 (0.5), 10.3 (0.9) at 19–21 h for the normobaric, 203 kPa, 243 kPa and 284 kPa conditions respectively. There was no significant association between flow rate and time period (P = 0.166) but the 203 kPa flow rates were significantly faster than the other flow rates (P = 0.008). In retrospect, the 203 kPa experiments had all been conducted with the same antibiotic solution (ceftazidime 6 g). Repeating that experimental arm using flucloxacillin 8 g produced flow rates of 10.4 (0.8) ml∙h-1, with no significant associations between flow rate and time period (P = 0.652) or pressure (P = 0.705). Conclusion: In this study, the flow rate of the Baxter LV10 InfusorTM device was not significantly affected by increases in ambient pressure across the pressure range of 101.3 kPa to 284 kPa, and flow rates were generally within a clinically acceptable range of 9−12 ml∙h-1. However, there was evidence that the specific antibiotic solution might affect flow rates and this requires further study.Keywords
This publication has 9 references indexed in Scilit:
- Performance of the Baxter Infusor LV10 under hyperbaric conditions.2015
- Hyperbaric intensive care technology and equipment.2015
- Medical devices and procedures in the hyperbaric chamber.2014
- Effects of atmospheric pressure conditions on flow rate of an elastomeric infusion pumpAmerican Journal of Health-System Pharmacy, 2012
- Mechanical performances of elastomers used in diffusersMedical Devices: Evidence and Research, 2011
- The Clinical and Economic Potential of Hyperbaric Oxygen Therapy in the Treatment of Diabetic Ulceration and Other ConditionsThe International Journal of Lower Extremity Wounds, 2007
- Performance of Infusion Pumps during Hyperbaric ConditionsAnesthesiology, 2002
- Failure of a Patient-controlled Analgesia Pump in a Hyperbaric EnvironmentAnesthesiology, 1999
- Hyperbaric-Oxygen TherapyNew England Journal of Medicine, 1996