Abstract
The mass flow and chemical kinetic equations for a gaseous chemiluminescence (CL) analyzer are formulated and solved. The resultant equation can be used to predict the absolute response of the analyzer as a function of the sample flow rate, the sample gas pressure, the chamber pressure, the chamber volume, the mass flow rate and mole fraction of the reagent gas, and the rate constants of the relevant chemical processes. Thus, the equation allows optimization of these parameters. It is shown that for varying sample pressure the analyzer can be used to measure either concentration or mole fraction and that interfering reactions can sometimes be discriminated against by chamber pressure variation. The equations apply equally well to a flowing‐liquid‐phase CL analyzer, if the chemical mechanism considered is appropriate.Keywords
This publication has 22 references indexed in Scilit:
- Optimization of response of chemiluminescence analyzers under various operating conditionsAnalytical Chemistry, 1983
- Monitor for measuring the total concentration of reactive hydrocarbons in ambient air based on their chemiluminescence reaction with oxygen atomsAnalytica Chimica Acta, 1981
- Measurements of nitric oxide in the stratosphere of the Southern HemisphereQuarterly Journal of the Royal Meteorological Society, 1980
- Nitrogen oxides (NOx) (= nitrogen oxide (NO) + nitrogen dioxide) monitor based on an hydrogen-atom direct chemiluminescence methodEnvironmental Science & Technology, 1980
- Detection of vinyl chloride and related compounds by a gas chromatographic, chemiluminescence techniqueEnvironmental Science & Technology, 1976
- Homogeneous gas phase chemiluminescence measurement of reactive hydrocarbon air pollutants by reaction with oxygen atomsEnvironmental Science & Technology, 1975
- Optimization of the operating parameters of chemiluminescent nitric oxide detectorsAnalytical Chemistry, 1974
- Low-pressure gas-phase ozone-olefin reactions. Chemiluminescence, kinetics, and mechanismsJournal of the American Chemical Society, 1974
- An instrument for nitric oxide measurements in the stratosphereReview of Scientific Instruments, 1974
- Analysis of Ozone and Nitric Oxide by a Chemiluminescent Method in Laboratory and Atmospheric Studies of Photochemical SmogJournal of the Air Pollution Control Association, 1972