Abstract
The ability to obtain the accurate mass of a protein in a complex sample mixture aids in determining its correct in vivo form. This is important when identifying post‐translationally modified proteins, protein variants or isoforms. The central technique used to separate proteins, 2‐dimensional gel electrophoresis offers excellent separation capabilities but does not provide adequate mass accuracy. In this study, an alternative method, liquid chromatography (LC) coupled with matrix‐assisted laser desorption/ionization‐time of flight (MALDI‐TOF)‐MS (LC‐MALDI) is described. LC‐MALDI‐MS was used to separate and determine the mass of proteins and peptides in a complex biological sample (i.e., human pituitary gland homogenate). Peptides and proteins were first separated by capillary chromatography and the eluent mixed post‐column with sinapinic acid matrix. The flow was then deposited directly onto a standard MALDI target via a capillary nebulizer. In addition to offering high mass accuracy, this method can be applied to peptide and protein quantification.

This publication has 31 references indexed in Scilit: