Generalized method for seeking q-preserving configurations of multi-pass cells
- 30 April 2019
- journal article
- research article
- Published by Optica Publishing Group in Optics Express
- Vol. 27 (10) , 14054-14063
Abstract
A method is presented for seeking specified multi-pass cell (MPC) configurations with q-parameters of the input and output Gaussian beams that remain the same (q-preserving configurations). The Frobenius norm (F-norm) is chosen as the quantitative criterion of the deviation between the transfer matrix of the MPC configurations and identity matrix. When the deviation is down near to zero, the corresponding configuration is close to an ideal q-preserving configuration. In contrast to common 2 × 2 transfer matrixes that are only applicable to MPCs with negligible astigmatism, we adopt 4 × 4 transfer matrixes that are also workable for multi-pass systems with astigmatism. To demonstrate the validation of the method, several q-preserving structures of four-objective multi-pass matrix system (FO-MMS) and double-row circular MPC (DR-CMPC), which are two fundamentally different MPC types, are illustrated. The variation tendencies of the quantitative deviation between the transfer matrix and the identity matrix are discussed in detail for each structure as functions of structural parameters. The optimal q-preserving structures are found for FO-MMS and DR-CMPC, with the matrix deviation of 0.0047 and 0.0051, while the corresponding total optical path length (OPL) are 33.6 m and 67.8 m, respectively. The OPL of the optimal DR-CPMC is longer than the traditional CMPC in the state of the art, and the deviation is three orders of magnitude smaller if the similar spherical reflecting surfaces are adopted.Keywords
Funding Information
- National key research and development program-Development Project of major scientific instruments and equipment (2018YFF0109600)
This publication has 20 references indexed in Scilit:
- Mid-infrared spectroscopy for gases and liquids based on quantum cascade technologiesThe Analyst, 2013
- Versatile multipass cell for laser spectroscopic trace gas analysisApplied Physics B, 2012
- A simple yet effective multipass reflector for vibrational excitation in molecular beamsReview of Scientific Instruments, 2008
- New generation of multipass systems in high resolution spectroscopySpectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 1996
- Astigmatic mirror multipass absorption cells for long-path-length spectroscopyApplied Optics, 1995
- Optical multipass matrix systemsApplied Optics, 1991
- A New White Type Multiple Pass Absorption CellApplied Optics, 1970
- Folded Optical Delay LinesApplied Optics, 1965
- Off-Axis Paths in Spherical Mirror InterferometersApplied Optics, 1964
- Rotation-Vibration Spectra of Diatomic and Simple Polyatomic Molecules with Long Absorbing Paths. I. The Spectrum of Fluoroform (CHF3) from 2.4μ to 0.7μThe Journal of Chemical Physics, 1948