0.4 μm spatial resolution with 1 GHz (λ=30 cm) evanescent microwave probe
Authors:
- 1 March 1999
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (3) , 1725-1729
Abstract
In this article we describe evanescent field imaging of material nonuniformities with a record resolution of 0.4 μm at 1 GHz (λg/750 000), using a resonant stripline scanning microwave probe. A chemically etched tip is used as a point-like evanescent field emitter and a probe–sample distance modulation is employed to improve the signal-to-noise ratio. Images obtained by evanescent microwave probe, by optical microscope, and by scanning tunneling microscope are presented for comparison. Probe was calibrated to perform quantitative conductivity measurements. The principal factors affecting the ultimate resolution of evanescent microwave probe are also discussed.Keywords
This publication has 4 references indexed in Scilit:
- High spatial resolution quantitative microwave impedance microscopy by a scanning tip microwave near-field microscopeApplied Physics Letters, 1997
- Surface resistance imaging with a scanning near-field microwave microscopeApplied Physics Letters, 1997
- Non-destructive characterization of materials by evanescent microwavesMeasurement Science and Technology, 1993
- Super-resolution Aperture Scanning MicroscopeNature, 1972