Assessment of central corneal thickness in normal, keratoconus, and post-laser in situ keratomileusis eyes using Scheimpflug imaging, spectral domain optical coherence tomography, and ultrasound pachymetry
- 1 June 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 36 (6) , 954-964
Abstract
To compare the central corneal thickness (CCT) in normal eyes, eyes with keratoconus, and eyes after laser in situ keratomileusis (LASIK) using 3 methods. Cornea Clinic, Grewal Eye Institute, Chandigarh, India. In this study, CCT was measured by sequential Scheimpflug imaging, spectral-domain anterior segment optical coherence tomography (AS-OCT), and ultrasound (US) pachymetry. Each of the 3 groups comprised 50 eyes. There were no differences between the 3 groups in age, sex, or intraocular pressure. In normal eyes, CCT was statistically significantly higher by US pachymetry (mean 525.8 microm +/- 41.4) [SD] than by Scheimpflug imaging (mean 519.4 +/- 40.9 microm) and AS-OCT (mean 517.9 +/- 41.5 microm) (both P<.001). In keratoconus eyes, CCT by US pachymetry (mean 446.4 +/- 57.9 microm) was statistically significantly higher than by Scheimpflug imaging (mean 439.6 +/- 62.1 microm) (P = .002) and AS-OCT (mean 441.8 +/- 58.4 microm) (P = .007). In post-LASIK eyes CCT by US pachymetry (mean 462.4 +/- 44.7 microm) was significantly higher than by Scheimpflug imaging (mean 457.9 +/- 33.6 microm) (P = .012) and AS-OCT (mean 455.4 +/- 43.2 microm) (P<.001). In all groups, CCT measured by Scheimpflug imaging and AS-OCT was similar. There was a statistically significant difference between Scheimpflug imaging and US pachymetry and AS-OCT, with US pachymetry measurements being consistently thicker. Thus, CCT should be interpreted in the context of the instrument used.Keywords
This publication has 34 references indexed in Scilit:
- Reproducibility and Repeatability of Central Corneal Thickness Measurement in Keratoconus Using the Rotating Scheimpflug Camera and Ultrasound PachymetryAmerican Journal of Ophthalmology, 2007
- Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopiaJournal of Cataract & Refractive Surgery, 2007
- Comparison of Central Corneal Thickness Measurements by Rotating Scheimpflug Camera, Ultrasonic Pachymetry, and Scanning-Slit Corneal TopographyOphthalmology, 2006
- Comparison of Three Methods of Measuring Corneal Thickness and Anterior Chamber DepthAmerican Journal of Ophthalmology, 2006
- Management of keratoconus with IntacsAmerican Journal of Ophthalmology, 2003
- Human Corneal Thickness and Its Impact on Intraocular Pressure MeasuresSurvey of Ophthalmology, 2000
- Deep Lamellar Keratoplasty Using Viscoelastic DissectionArchives of Ophthalmology (1950), 1999
- Live-epikeratophakia for keratoconusJournal of Cataract & Refractive Surgery, 1998
- Accuracy of ultrasonic pachymetry and videokeratography in detecting keratoconusJournal of Cataract & Refractive Surgery, 1998
- KeratoconusSurvey of Ophthalmology, 1998