Developing a Small Animal Model of Extracorporeal Circulation
Authors:
- Judiann MiskulinJMJudiann Miskulin
- Robert H. BartlettRBRobert H. Bartlett
- Gail M. AnnichGAGail M. Annich
- 1 May 2006
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 52 (3) , 296-302
Abstract
To identify nonthrombogenic devices to be used in extracorporeal circulation (ECC), an efficient, small animal model is required. Initially, a venovenous (VV) model in rabbits was designed for this purpose and was a good representation of ECC. Technical difficulties in the VV model led to the development of a more simplistic arteriovenous (AV) model. Anesthetized, tracheotomized, 3-kg rabbits were used for both models. Circuits were constructed of PVC tubing. The VV model used 8-Fr umbilical artery catheters for both drainage and reinfusion, and the AV model used a 14-GA angiocatheter for carotid artery access and a 10-Fr thoracic catheter for venous access. The AV model included a chamber to mimic oxygenator or filter modeling. Hourly measurements included blood gases, platelet counts, and fibrinogen levels for the 4-hour studies. The VV ECC groups demonstrated platelet consumption like that seen in the clinical arena. The AV model demonstrated the same with or without additional surface area within the chamber. The AV model was deemed to be superior due to its simplicity, ability for filter modeling, and decrease in intensive monitoring. However, both models are excellent designs for nonthrombogenic surface testing.Keywords
This publication has 12 references indexed in Scilit:
- Coated versus Noncoated Circuits in Pediatric Cardiopulmonary BypassASAIO Journal, 2005
- Biomaterial-associated thrombosis: roles of coagulation factors, complement, platelets and leukocytesBiomaterials, 2004
- Reduced platelet activation and thrombosis in extracorporeal circuits coated with nitric oxide release polymersCritical Care Medicine, 2000
- Thromboresistant Chemical Sensors Using Combined Nitric Oxide Release/Ion Sensing Polymeric FilmsJournal of the American Chemical Society, 1997
- Contact Activation, Heparins and Cardiopulmonary BypassThrombosis and Haemostasis, 1992
- Platelet consumption during neonatal extracorporeal life support (ECLS)Perfusion, 1992
- Review article : Blood compatibility of cardiopulmonary bypass circuitsPerfusion, 1987
- Survival of density subpopulations of rabbit platelets: use of 51Cr-or 111In-labeled platelets to measure survival of least dense and most dense platelets concurrentlyBlood, 1983
- Indium-111 labeled platelets: Studies on preparation and evaluation of in vitro and in vivo functionsThrombosis Research, 1976
- Adenosine Diphosphate‐induced Platelet Aggregation in Suspensions of Washed Rabbit PlateletsBritish Journal of Haematology, 1970