Cooled Needle Catheter Ablation Creates Deeper and Wider Lesions Than Irrigated Tip Catheter Ablation
- 12 April 2005
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 16 (5) , 508-515
Abstract
Objectives: To design and test a catheter that could create deeper ablation lesions. Background: Endocardial radiofrequency (RF) ablation is unable to reliably create transmural ventricular lesions. We designed an intramural needle ablation catheter with an internally cooled 1.1‐mm diameter straight needle that could be advanced up to 14 mm into the myocardium. The prototype catheter was compared with an irrigated tip ablation catheter. Methods: Ablation lesions were created under general anesthesia in 14 male sheep (weight 44 ± 7.3 kg) with fluoroscopic guidance. Each of the catheters was used to create two ablation lesions at randomly allocated positions within the left ventricle. The irrigation rate, target temperature, and maximum power were: 20 mL/min, 85°C, 50 W for the intramural needle catheter and 20 mL/min, 50°C, 50 W for the irrigated tip catheter, respectively. All ablations were performed for 2 minutes. After the last ablation, blue tetrazolium (12.5 mg/kg) was infused intravenously. The heart was removed via a left thoracotomy after monitoring the sheep for one hour. Results: There was no evidence of cardiac tamponade in any sheep. The intramural needle catheter lesions were significantly wider (10.9 ± 2.8 mm vs 10.1 ± 2.4 mm, P = 0.01), deeper (9.6 ± 2.0 mm vs 7.0 ± 1.3 mm, P = 0.01), and more likely to be transmural (38% vs 0%, P = 0.03). Conclusions: Cooled intramural needle ablation creates lesions that are significantly deeper and wider than endocardial RF ablation using an irrigated tip catheter in sheep hearts. This technology may be useful in treating ventricular tachycardia resistant to conventional ablation techniques.Keywords
This publication has 18 references indexed in Scilit:
- Cooled Intramural Needle Catheter Ablation Creates Deeper Lesions than Irrigated Tip Catheter AblationPacing and Clinical Electrophysiology, 2004
- Intramural Radiofrequency Ablation:Pacing and Clinical Electrophysiology, 2004
- Deep Myocardial Ablation Lesions Can Be Created with a Retractable Needle‐Tipped CatheterPacing and Clinical Electrophysiology, 2004
- A new catheter design using needle electrode for subendocardial RF ablation of ventricular muscles: finite element analysis and in vitro experimentsIEEE Transactions on Biomedical Engineering, 2000
- Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine MyocardiumCirculation, 1999
- Radiofrequency Delivery Through a Cooled Catheter Tip Allows the Creation of Larger Endomyocardial Lesions in the Ovine HeartJournal of Cardiovascular Electrophysiology, 1995
- Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle PreparationCirculation, 1995
- Development of a new intraoperative radiofrequency ablation technique using a needle electrodeThe Annals of Thoracic Surgery, 1994
- Functional role of the epicardium in postinfarction ventricular tachycardia. Observations derived from computerized epicardial activation mapping, entrainment, and epicardial laser photoablation.Circulation, 1991
- The Effect of Electrode Design on the Efficiency of Delivery of Radiofrequency Energy to Cardiac Tissue In VitroPacing and Clinical Electrophysiology, 1989