Directional Transverse Oscillation Vector Flow Estimation
Open Access
- 1 June 2017
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 64 (8) , 1194-1204
Abstract
A method for estimating vector velocities using transverse oscillation (TO) combined with directional beamforming is presented. In directional TO (DTO), a normal focused field is emitted and the received signals are beamformed in the lateral direction transverse to the ultrasound beam to increase the amount of data for vector velocity estimation. The approach is self-calibrating as the lateral oscillation period is estimated from the directional signal through a Fourier transform to yield quantitative velocity results over a large range of depths. The approach was extensively simulated using Field IIpro and implemented on the experimental Synthetic Aperture Real-time Ultrasound System (SARUS) scanner in connection with a BK Medical 8820e convex array transducer. Velocity estimates for DTO are found for beam-to-flow angles of 60°, 75°, and 90°, and vessel depths from 24 to 156 mm. Using 16 emissions, the standard deviation (SD) for angle estimation at depths ranging from 24 to 104 mm is between 6.01° and 0.93° with a mean SD of 2.8°. The mean relative SD for the lateral velocity component is 9.2% and the mean relative bias -3.4% or four times lower than for traditional TO. The approach also works for deeper lying vessels with a slight increase in SD to 15.7%, but a maintained bias of -3.5% from 126 to 156 mm. Data for a pulsating flow have also been acquired for 15 cardiac cycles using a CompuFlow 1000 pump. The relative SD was here 7.4% for a femoral artery waveform.Keywords
Funding Information
- Danish Advanced Technology Foundation (82-2012-4)
- BK Ultrasound ApS, Herlev, Denmark
This publication has 38 references indexed in Scilit:
- A multi-threaded version of Field II2014
- SARUS: A synthetic aperture real-time ultrasound systemIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2013
- A Real-Time 2-D Vector Doppler System for Clinical ExperimentationIEEE Transactions on Medical Imaging, 2008
- Estimation of velocity vector angles using the directional cross-correlation methodIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2006
- Estimation of Velocity Vectors in Synthetic Aperture Ultrasound ImagingIEEE Transactions on Medical Imaging, 2006
- Experimental evaluation of velocity and power estimation for ultrasound blood flow imaging, by means of a two-dimensional autocorrelation approachIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1995
- An axial velocity estimator for ultrasound blood flow imaging, based on a full evaluation of the Doppler equation by means of a two-dimensional autocorrelation approachIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1995
- Multifrequency Doppler: improving the quality of spectral estimation by making full use of the information present in the backscattered RF echoesIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1994
- Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducersIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1992
- Real-Time Two-Dimensional Blood Flow Imaging Using an Autocorrelation Technique1985