Successful application of snare‐kissing‐catheter technique to implant leadless pacemaker in severely dilated right heart
- 25 January 2021
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 44 (5) , 960-962
Abstract
Implantation of leadless pacemaker is efficacy and safety compared with the traditional pacemaker in structurally normal hearts. However, delivery experience of leadless pacemaker in patients with severe right heart enlargement remains limited. We present the rare case of a patient with giant right heart and moderate to severe tricuspid regurgitation implanted with a leadless Micra transcatheter pacemaker system. The extension of the Micra delivery catheter can be improved by using a single‐loop snare on the catheter proximal to appropriate right ventricle (RV) pacing position. The snare‐kissing‐catheter technique can aid in successful deployment in the setting of challenging right heart enlargement.Keywords
This publication has 7 references indexed in Scilit:
- Use of a snare to aid implanting leadless pacemaker in a patient with severely dilated atriaEP Europace, 2020
- Nanostim leadless pacemaker retrieval and simultaneous micra leadless pacemaker replacement: a single-center experienceJournal of Interventional Cardiac Electrophysiology, 2019
- Successful leadless pacemaker deployment in a patient with challenging right heart anatomy using a double snare techniqueHeartRhythm Case Reports, 2019
- Leadless pacemaker: Performance and complicationsTrends in Cardiovascular Medicine, 2018
- Monocentric experience of leadless pacing with focus on challenging cases for conventional pacemakerActa Cardiologica, 2017
- A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval RegistryHeart Rhythm, 2017
- Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing StudyEuropean Heart Journal, 2015