Comparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometry
Authors:
- Jinhai HuangJHJinhai Huang
- Konrad PesudovsKPKonrad Pesudovs
- Daizong WenDWDaizong Wen
- Shihao ChenSCShihao Chen
- Thomas WrightTWThomas Wright
- Xiaoyu WangXWXiaoyu Wang
- Yini LiYLYini Li
- Qinmei WangQWQinmei Wang
- 1 February 2011
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 37 (2) , 341-348
Abstract
To compare central corneal thickness (CCT), anterior chamber depth (ACD), and keratometry (K) readings measured using optical low-coherence reflectometry (OLCR) biometry and high-resolution rotating Scheimpflug photography. Eye Hospital of Wenzhou Medical College, Wenzhou, China. Comparative case series. The CCT, ACD endothelium to lens, ACD epithelium to lens, and K (mean; in flattest meridian; in steepest meridian) were measured 5 times using the LenStar/Biograph OLCR biometer and 3 times with the Pentacam Scheimpflug system in eyes of healthy volunteers. Concordance was evaluated using paired t tests, the Pearson correlation, and Bland-Altman analyses. The CCT, ACD endothelium to lens, and ACD epithelium to lens measured with the Scheimpflug system were slightly, albeit significantly, higher than with the OLCR biometer (P<.05); the respective 95% limits of agreement (LoA) were -8.2 μm to 15.7 μm, -0.11 to 0.15 mm, and -0.13 to 0.17 mm. However, the Scheimpflug system gave significantly flatter readings for K in the flattest meridian (95% LoA, -0.54 to 0.32 diopters [D]), K in the steepest meridian (95% LoA, -0.63 to 0.45 D), and mean K (95% LoA, -0.53 to 0.33 D) (P<.001). The CCT, ACD, and K readings were all highly correlated between the 2 devices (r >0.95, P<.001). The CCT and ACD measurements with the OLCR biometer and Scheimpflug system can be used interchangeably in healthy young subjects. However, for K measurements, these devices have wide LoA so may not be interchangeable under certain clinical circumstances.Keywords
This publication has 25 references indexed in Scilit:
- Comparison of anterior and posterior elevation map measurements between 2 Scheimpflug imaging systemsJournal of Cataract & Refractive Surgery, 2009
- Contralateral Eye Study of Corneal Collagen Cross-linking With Riboflavin and UVA Irradiation in Patients With KeratoconusJournal of Refractive Surgery, 2009
- Accuracy of a new partial coherence interferometry analyser for biometric measurementsBritish Journal of Ophthalmology, 2009
- Long-term Follow-up of Endothelial Cell Change after Artisan Phakic Intraocular Lens ImplantationOphthalmology, 2008
- Biometry and intraocular lens power calculationCurrent Opinion in Ophthalmology, 2008
- Risk Assessment for Ectasia after Corneal Refractive SurgeryOphthalmology, 2008
- Anterior segment biometry with the Pentacam: Comprehensive assessment of repeatability of automated measurementsJournal of Cataract & Refractive Surgery, 2008
- Anterior segment changes with age and during accommodation measured with partial coherence interferometryJournal of Cataract & Refractive Surgery, 2007
- Flap and stromal bed thickness in laser in situ keratomileusis enhancementJournal of Cataract & Refractive Surgery, 2004
- Measurement of accommodation after implantation of an accommodating posterior chamber intraocular lensJournal of Cataract & Refractive Surgery, 2003