Abstract
PURPOSE: To determine the effects of cyst diameter and location (intrarenal, exophytic), renal attenuation, section collimation, and computed tomographic (CT) interscanner variability on renal cyst pseudoenhancement in a phantom model. MATERIALS AND METHODS: A customized anthropomorphic phantom was designed to accept 40-, 140-, and 240-HU renal inserts containing intrarenal and exophytic 7-, 10-, and 15-mm cysts. Each phantom and insert were scanned with five different helical CT scanners by using 1.0–1.5-mm, 2.50–3.75-mm, 5.0-mm, 7.0–8.0-mm, and 10.0-mm section collimation. Means and SDs of CT number measurements were obtained for each cyst within each variably “enhanced” renal insert. Mixed-model analysis of variance accommodating heteroscedasticity of data was used to assess the effect of scanner type, section collimation, and cyst diameter on cyst attenuation. RESULTS: Pseudoenhancement (range, 10.3–28.3 HU), observed by using effective section collimation equal to or less than 50% of cyst diameter, o...