A technique of double‐resonant operation of 19F and 1H quadrature birdcage coils
Authors:
- 1 May 1991
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 19 (1) , 180-185
Abstract
A technique was developed to change the resonant frequency of a rotating field birdcage coil from 19F resonance (76.14 MHz) to 1H resonance (80.92 MHz). This was done by inserting an additional shield with a smaller radius on the coil which resonates at 19F. It was shown that the inhomogeneity introduced by reducing the shield radius is not significant. A working coil built with this technique produced both 19F and 1H images satisfactorily. © 1991 Academic Press, Inc.Keywords
This publication has 6 references indexed in Scilit:
- A novel rotating-field coil for sodium imagingJournal of Magnetic Resonance (1969), 1991
- Hydrogen/fluorine retuning tomography. applications to 1h image‐guided volume‐selective 19f spectroscopy and relaxometry of perfluorocarbon emulsions in tissueMagnetic Resonance in Medicine, 1990
- A technique for double resonant operation of birdcage imaging coilsIEEE Transactions on Medical Imaging, 1989
- A multiple-frequency coil with a highly uniform B1 fieldJournal of Magnetic Resonance (1969), 1989
- Comparison of linear and circular polarization for magnetic resonance imagingJournal of Magnetic Resonance (1969), 1985
- An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5 TJournal of Magnetic Resonance (1969), 1985