Comparability of a Blood-Pressure-Monitoring Smartphone Application with Conventional Measurements—A Pilot Study
Authors:
- Annina S. VischerAnnina S. Vischer
- Jana RosaniaJana Rosania
- Thenral SocratesTSThenral Socrates
- Christina BlaschkeCBChristina Blaschke
- Jens EcksteinJEJens Eckstein
- Yara-Maria ProustYPYara-Maria Proust
- Guillaume BonnierGBGuillaume Bonnier
- Martin ProençaMPMartin Proença
- Mathieu LemayMLMathieu Lemay
- Thilo BurkardTBThilo Burkard
Open Access
- 19 March 2022
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 12 (3) , 749
Abstract
(1) Background: New cuffless technologies attempting blood-pressure measurements (BPM) offer possibilities to improve hypertension awareness and control. The aim of this study was to compare a smartphone application (app)-based algorithm with office BPM (OBPM). (2) Methods: We included consecutive patients with an indication for ambulatory BPM. The smartphone app (RIVA digital) acquired the pulse wave in the fingers’ arterial bed using the phone’s camera and estimated BP based on photoplethysmographic (PPG) waveforms. Measurements were alternatingly taken with an oscillometric cuff-based device and smartphone BPM (AppBP) on two consecutive days. AppBP were calibrated to the first OBPM. Each AppBP was compared to its CuffBP (mean of the previous/following OBPM). (3) Results: 50 participants were included, resulting in 50 AppBP values on Day 1 and 33 on Day 2 after exclusion of 225 AppBP due to insufficient quality. The mean ± SD of the differences between AppBP and CuffBP was 0.7 ± 9.4/1.0 ± 4.5 mmHg (p-value 0.739/0.201) on Day 1 and 2.6 ± 8.2/1.3 ± 4.1 mmHg (p-value 0.106/0.091) on Day 2 for systolic/diastolic values, respectively. There were no significant differences between the deviations on Day 1 and Day 2 (p-value 0.297/0.533 for systolic/diastolic values). Overall, there were 10 (12%) systolic measurement pairs differing by >15 mmHg. (4) Conclusions: In this pilot evaluation, the RIVA Digital app shows promising results when compared to oscillometric cuff-based measurements, especially regarding diastolic values. Its differences between AppBP–CuffBP have a good stability one day after calibration. Before clinical use, signal acquisition needs improvement and the algorithm needs to undergo formal validation against a gold-standard BPM method.Keywords
Funding Information
- Swiss Center for Electronics and Microtechnology (Switzerland) (NA)
This publication has 36 references indexed in Scilit:
- Prevalence, awareness, treatment, and control of hypertension in China: data from 1·7 million adults in a population-based screening study (China PEACE Million Persons Project)The Lancet, 2017
- Validation of the Instant Blood Pressure Smartphone AppJAMA Internal Medicine, 2016
- Blood pressure lowering for prevention of cardiovascular disease and death: a systematic review and meta-analysisThe Lancet, 2016
- Hypertension prevalence, awareness, treatment and control in Germany 1998 and 2008–11Journal of Human Hypertension, 2014
- Validation of the Welch Allyn SureBP (inflation) and StepBP (deflation) algorithms by AAMI standard testing and BHS data analysisBlood Pressure Monitoring, 2011
- Does self-monitoring reduce blood pressure? Meta-analysis with meta-regression of randomized controlled trialsAnnals of Medicine, 2010
- European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adultsBlood Pressure Monitoring, 2010
- Age-specific relevance of usual blood pressure to vascular mortality: a meta-analysis of individual data for one million adults in 61 prospective studiesThe Lancet, 2002
- The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systemsJournal Of Hypertension, 1990
- Increased Absenteeism from Work after Detection and Labeling of Hypertensive PatientsNew England Journal of Medicine, 1978