Detection of a coronary artery vessel wall: performance of 0.3 mm fine-cell detector computed tomography—a phantom study

Authors:
  • MJMasahiro Jinzaki
    Masahiro Jinzaki
  • YTYutaka Tanami
    Yutaka Tanami
  • KMKazuhiro Matsumoto
    Kazuhiro Matsumoto
  • AUAkihisa Ueno
    Akihisa Ueno
  • MNMasatake Nukui
    Masatake Nukui
  • YIYasuhiro Imai
    Yasuhiro Imai
  • YIYotaro Ishihara
    Yotaro Ishihara
  • ANAkihiko Nishide
    Akihiko Nishide
  • KSKosuke Sasaki
    Kosuke Sasaki
  • SKSachio Kuribayashi
    Sachio Kuribayashi

Abstract
The purpose of this study was to evaluate whether experimental fine-cell detector computed tomography with a 0.3125 mm cell (0.3 mm cell CT) can improve the detection of coronary vessel walls compared with conventional 64-slice computed tomography with a 0.625 mm cell (0.6 mm cell CT). A coronary vessel wall phantom was scanned using 0.6 mm cell CT and 0.3 mm cell CT. The data for 0.3 mm cell CT were obtained using four protocols: a radiation dose equal, double, triple or quadruple that were used in the 0.6 mm cell CT protocol. The detectable size of the vessel wall was assessed based on the first and second derivative functions, and the minimum measurable values were compared using a paired t-test. As a result, the minimum detectable wall thickness of 0.6 mm cell CT (1.5 mm) was significantly larger than that of 0.3 mm cell CT performed using the triple- and quadruple-dose protocols (0.9 mm) and the double-dose protocol (1.1 mm). The difference in the minimum detectable vessel wall thickness measured using 0.6 mm cell CT (1.5 ± 0.1 mm) and 0.3 mm cell CT (0.9 ± 0.1 mm, 1.1 ± 0.2 mm) was significant (p < 0.01). We concluded that 0.3 mm cell CT improved the detection of coronary vessel walls when a more than double-dose protocol was used compared with 0.6 mm cell CT.

This publication has 33 references indexed in Scilit: