Abstract
In order to enhance 3D image data from magnetic resonance angiography (MRA), a novel method based on the theory of multidimensional adaptive filtering has been developed. The purpose of the technique is to suppress image noise while enhancing important structures. The method is based on local structure estimation using six 3D orientation selective filters, followed by an adaptive filtering step controlled by the local structure information. The complete filtering procedure requires approximately 3 minutes of computational time on a standard workstation for a 256 × 256 × 64 data set. The method has been evaluated using a mathematical vessel model and in vivo MRA data (both phase contrast and time of flight (TOF)). 3D adaptive filtering results in a better delineation of small blood vessels and efficiently reduces the high‐frequency noise. Depending on the data acquisition and the original data type, contrast‐to‐noise ratio (CNR) improvements of up to 179% (8.9 dB) were observed. 3D adaptive filtering may provide an alternative to prolonging the scan time or using contrast agents in MRA when the CNR is low. J. Magn. Reson. Imaging 2001;14:63–71.Keywords
Funding Information
- Swedish Medical Science Foundation (B94-39X-10885-01)
- NIH (P41-RR13218, R01-RR11747;)
- Center for Integration of Medicine and Innovative Technology (CIMIT)
This publication has 27 references indexed in Scilit:
- Directional low-pass filtering for improved accuracy and reproducibility of stenosis quantification in coronary arteriogramsIEEE Transactions on Medical Imaging, 1995
- Mathematical basis of eigenimage filteringMagnetic Resonance in Medicine, 1994
- Generalized matched filtering for time‐resolved MR angiography of pulsatile flowMagnetic Resonance in Medicine, 1993
- Noise reduction in three‐dimensional phase‐contrast MR velocity measurementslJournal of Magnetic Resonance Imaging, 1993
- Improvements of nuclear magnetic resonance image quality using iterations of adaptive nonlinear filteringIEEE Transactions on Medical Imaging, 1993
- A weighted least‐squares method for nuclear magnetic resonance velocity imagingMagnetic Resonance in Medicine, 1993
- Nonlinear anisotropic filtering of MRI dataIEEE Transactions on Medical Imaging, 1992
- A comparative analysis of several transformations for enhancement and segmentation of magnetic resonance image scene sequencesIEEE Transactions on Medical Imaging, 1992
- Displays for MR angiographyMagnetic Resonance in Medicine, 1990
- Anisotropic Nonstationary Image Estimation and Its Applications: Part I--Restoration of Noisy ImagesIEEE Transactions on Communications, 1983