Technical note: Energy dependence of the Gafchromic EBT4 film: Dose‐response curves for 70 kV, 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV x‐ray beams
Authors:
- 1 February 2023
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 50 (6) , 3738-3745
Abstract
Background: EBT4 was newly released for radiotherapy quality assurance to improve the signal‐to‐noise ratio in radiochromic film dosimetry. It is important to know its dose‐response characteristics before its use in the clinic.Purpose: This study aims to investigate and compare the dose‐response curves of the Gafchromic EBT4 film for megavoltage and kilovoltage x‐ray beams with different dose levels, scanning spatial resolutions, and sizes of region of interest (ROI).Methods: EBT4 film (Lot#07052201) calibration strips (3.5 × 20 cm2) were exposed to a 10×10 cm2 open field at doses of 0, 63, 125, 500, 750, 1000 cGy using 6 MV photon beam. EBT4 film strips from the same lot were then exposed to each x‐ray beam (6 MV, 6 MV FFF, 10 MV FFF, 15 MV, and 70 kV) at six dose values (50, 100, 300, 600, 800, 1000 cGy). A full sheet (25 × 20 cm2) of EBT4 film was irradiated at each energy with 300 cGy for profile comparison with the treatment planning calculation. At two different spatial resolutions of 72 and 300 dpi, each film piece was scanned three consecutive times in the center of an Epson 10000XL flatbed scanner in 48‐bit color. The scanned images were analyzed using FilmQA Pro. For each scanned image, an ROI of 2 × 2 cm2 at the field center was selected to obtain the average pixel value with its standard deviation in the ROI. An additional ROI of 1 cm diameter circle was also used to evaluate the impact of ROI shape and size, especially for FFF beams. The dose value, average dose‐response value, and associated uncertainty were determined for each energy and relative responses were analyzed. The Student's t‐test was performed to evaluate the statistical significance of the dose‐response values with different color channels, ROI shapes, and spatial resolutions.Results: The dose‐response curves for the five x‐ray energies were compared in three color channels. Weak energy dependence was found among the megavoltage beams. No significant differences (average ∼1.1%) were observed for all doses in this study among 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV beams, regardless of spatial resolution and color channel. However, a statistically significant difference in dose‐response was observed up to 12% between 70 kV and 6 MV beams.Conclusions: The dose‐response curves for Gafchromic EBT4 films were nearly independent of the energy of the photon beams among 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV. For very low‐energy photons (e.g., 70 kV), a separate calibration from the same low‐energy x‐ray is necessary.Keywords
This publication has 27 references indexed in Scilit:
- The response of the new MD-V2-55 radiochromic film exposed to60Co gamma raysPhysics in Medicine & Biology, 2010
- Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energiesMedical Physics, 2010
- Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapyPhysics in Medicine & Biology, 2010
- Technical Note: Homogeneity of Gafchromic®EBT2 filmMedical Physics, 2010
- Monte Carlo calculated absorbed‐dose energy dependence of EBT and EBT2 filmMedical Physics, 2010
- Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetryPhysics in Medicine & Biology, 2009
- Analysis of uncertainties in Gafchromic® EBT film dosimetry of photon beamsPhysics in Medicine & Biology, 2008
- Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurancePhysics in Medicine & Biology, 2007
- Absorption spectra variations of EBT radiochromic film from radiation exposurePhysics in Medicine & Biology, 2005
- The use of a reflective scanner to study radiochromic film responsePhysics in Medicine & Biology, 2002