The effects of flattening filter‐free beams and aperture shape controller on the complexity of conventional large‐field treatment plans
Authors:
Open Access
- 1 August 2023
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 24 (11) , e14108
Abstract
Purpose: The purpose of this study was to investigate the impact of using flattening filter‐free (FFF) beams and the aperture shape controller (ASC) on the complexity of conventional large‐field treatment plans.Methods and materials: A total of 24 head and neck (H&N) and 24 prostate with pelvic nodes treatment plans were used in this study. Each plan was reoptimized using the original clinical objectives with both flattened and FFF beams, as well as six different ASC settings. The dosimetric qualities of each plan cohort were evaluated using commonly used dose‐volume histogram values, and plan complexities were assessed through metrics including monitor unit (MU)/Dose, change in gantry speed, multileaf collimator (MLC) speed, the edge area ratio metric (EM), and the equivalent square length.Results: No significant differences in dosimetric qualities were found between plans with flattened and FFF beams. The ASC settings did not have significant effects on dosimetric qualities in the H&N plan cohort, but the “very high” ASC setting resulted in poorer dosimetric results for the prostate plans. Plans with FFF beams had significantly higher MU/Dose compared to plans with flattened beams. The use of flattening filter (FF) had significant effects on the change in gantry speed, with flattened beams producing plans that required higher change in gantry speed. However, the FF did not have significant effects on MLC speed, EM, or equivalent square length. In contrast, ASC settings had significant effects on these three metrics; increasing the ASC level resulted in plans with decreasing MLC speed, lower edge area ratio, and higher equivalent square length.Conclusion: This study demonstrated that using FFF beams with various ASC settings, except for the “very high” level, can produce plans with reduced complexities without compromising dosimetric qualities in conventional large‐field treatment plans.Keywords
This publication has 36 references indexed in Scilit:
- Penalization of aperture complexity in inversely planned volumetric modulated arc therapyMedical Physics, 2012
- Volumetric modulated arc planning for lung stereotactic body radiotherapy using conventional and unflattened photon beams: a dosimetric comparison with 3D techniqueRadiation Oncology, 2011
- Feasibility and early clinical assessment of flattening filter free (FFF) based stereotactic body radiotherapy (SBRT) treatmentsRadiation Oncology, 2011
- Flattening filter free beams in SBRT and IMRT: Dosimetric assessment of peripheral dosesZeitschrift für Medizinische Physik, 2011
- Current status and future perspective of flattening filter free photon beamsMedical Physics, 2011
- Dosimetric Impact of Interplay Effect on RapidArc Lung Stereotactic Treatment DeliveryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Out-of-field photon dose following removal of the flattening filter from a medical acceleratorPhysics in Medicine & Biology, 2010
- Dosimetric characteristics of 6 and 10MV unflattened photon beamsRadiotherapy and Oncology, 2009
- Stereotactic radiotherapy for lung cancer using a flattening filter free ClinacJournal of Applied Clinical Medical Physics, 2009
- The Tradeoff Between Treatment Plan Quality and Required Number of Monitor Units in Intensity-modulated RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007