Development of x-ray imaging at the Daresbury SRS
Authors:
- 1 July 1989
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 60 (7) , 2464-2467
Abstract
A scanning transmission x-ray microscope has been used to form images with 380-eV x rays on the undulator beamline at the Daresbury synchrotron radiation source. A resolution of 80 nm has been achieved using Fresnel zone-plate focusing optics and images with reasonable signal-to-noise ratios can be acquired in less than 30 min.Keywords
This publication has 3 references indexed in Scilit:
- Contamination Lithography For The Fabrication Of Zone Plate X-Ray Lenses1985
- Soft x-ray microscopesReview of Scientific Instruments, 1985
- Transmission microscopy of unmodified biological materials. Comparative radiation dosages with electrons and ultrasoft X-ray photonsUltramicroscopy, 1976