pH measurement of low-conductivity waters
Authors:
- Eurybiades BusenbergEBEurybiades Busenberg
- Niel PlummerNPNiel Plummer
Open Access
- 1 January 1987
- report
- Published by US Geological Survey
Abstract
PH is an important and commonly measured parameter of precipitation and other natural waters. The various sources of errors in pH measurement were analyzed and procedures for improving the accuracy and precision of pH measurements in natural waters with conductivities of < 100 uS/cm at 25 C are suggested. Detailed procedures are given for the preparation of dilute sulfuric acid standards to evaluate the performance of pH electrodes in low conductivity waters. A daily check of the pH of dilute sulfuric acid standards and deionized water saturated with a gas mixture of low carbon dioxide at partial pressure (air) prior to the measurement of the pH of low conductivity waters is suggested. (Author 's abstract)Keywords
This publication has 10 references indexed in Scilit:
- DEVELOPMENT OF A STANDARD REFERENCE MATERIAL FOR RAINWATER ANALYSISJournal of Research of the National Bureau of Standards, 1986
- Improvements in the precision of pH measurements a laboratory reference electrode with renewable free-diffusion liquid junctionAnalytica Chimica Acta, 1985
- Evaluation of methods used for the determination of acidity in "acid rain" samples1985
- Methods for Reliable pH Measurements of Precipitation SamplesInternational Journal of Environmental Analytical Chemistry, 1984
- Effect of storage solution composition on pH combination electrode response times at low temperatureThe Analyst, 1984
- Procedures for the measurement of pH in low ionic strength solutions including freshwaterThe Analyst, 1983
- The solubilities of calcite, aragonite and vaterite in CO2-H2O solutions between 0 and 90°C, and an evaluation of the aqueous model for the system CaCO3-CO2-H2OGeochimica et Cosmochimica Acta, 1982
- Recent developments in ph standardisation and measurement for dilute aqueous solutionsAnalytica Chimica Acta, 1981
- A common source of error in pH measurementsBiochemical Journal, 1981
- Determination of pH: Theory and PracticeJournal of the Electrochemical Society, 1973