Objective The morphological changes of the pipeline embolization device (PED; Medtronic, Minneapolis, MN, USA), such as delayed migration or foreshortening, can relate to the incomplete occlusion of aneurysms. Case presentation A 30-year-old man with a giant cavernous carotid artery aneurysm was treated with two PEDs using the overlapping technique. Six months after treatment, follow-up angiography showed morphological changes of the PEDs and residual flow into the aneurysm. Chronological cone-beam computed tomography fusion imaging clearly revealed the dynamic foreshortening of the first PED and the disconnection of both PEDs, so we decided to implant an additional PED. Conclusion This case illustrates that a three-dimensional understanding can be useful for assessing the cause of treatment failure or recurrence.Keywords
This publication has 8 references indexed in Scilit:
- Metal artifact reduction algorithm for image quality improvement of cone-beam CT images of medium or large cerebral aneurysms treated with stent-assisted coil embolizationNeuroradiology, 2019
- Very delayed discontinuation of telescoped Pipeline embolization devices: a case reportInterventional Neuroradiology, 2018
- Safety and efficacy of the Pipeline embolization device for treatment of intracranial aneurysms: a pooled analysis of 3 large studiesJournal of Neurosurgery, 2017
- Learning from failure: persistence of aneurysms following pipeline embolizationJournal of Neurosurgery, 2017
- An outcomes-based grading scale for the evaluation of cerebral aneurysms treated with flow diversionJournal of NeuroInterventional Surgery, 2016
- Spontaneous Delayed Migration/Shortening of the Pipeline Embolization Device: Report of 5 CasesAmerican Journal of Neuroradiology, 2013
- Pipeline for Uncoilable or Failed Aneurysms: Results from a Multicenter Clinical TrialRadiology, 2013
- Endovascular Treatment of Intracranial Aneurysms With Flow DivertersStroke, 2013