Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study
Open Access
- 19 November 2020
- journal article
- research article
- Published by Springer Nature in Techniques in Coloproctology
- Vol. 25 (2) , 215-221
Abstract
Background: Sacral neuromodulation (SNM) is an established treatment option for patients with faecal incontinence. The location of the stimulating electrode is considered to be essential for treatment success. The purpose of this study was to evaluate the position of SNM electrodes after using a standardized fluoroscopy-guided implantation technique. Methods: For this cadaver study, SNM electrodes were implanted bilaterally in 5 lower body specimens. The lower edge of the sacroiliac joint and the medial edge of the sacral foramina were marked using fluoroscopy to draw an ‘H’ with the crossing points identifying S3. After electrode placement the pelvis was dissected to describe the exact position of the SNM electrodes. Results: The electrodes were inserted at an angle with a median degree measure of 60° (range 50–65°) to the skin, with a median distance of 9 mm (range 0–13 mm) from the S3 marking. All electrodes entered the third sacral foramen. The median distance of the electrodes to the sacral nerve was 0 mm (range 0–3 mm) for the most proximal, 0.5 mm (range 0–5 mm) for the second, 2.25 mm (range 0–11 mm) for the third and 1.75 mm (range 0–16 mm) for the most distant electrode. There was neither a significant difference in the proximity of the electrodes to the nerve between the right and left side (proximal to distal electrode: p = 0.18, p = 0.16, p = 0.07, p = 0.07) nor between male and female cadavers (p = 0.25, p = 0.21, p = 0.66, p = 0.66). Conclusions: A standardized fluoroscopy-guided implantation technique enables a close contact between electrode and nerve. This can potentially result in an improved clinical outcome.Keywords
Funding Information
- Medical University of Vienna
This publication has 13 references indexed in Scilit:
- Surface anatomical landmarks for the location of posterior sacral foramina in sacral nerve stimulationTechniques in Coloproctology, 2016
- Is the efficacy of sacral nerve stimulation for faecal incontinence dependent on the number of active electrode poles achieved during permanent lead insertion?Colorectal Disease, 2016
- Sacral nerve stimulation for faecal incontinence and constipation: a European consensus statementColorectal Disease, 2015
- Surface and boney landmarks for sacral neuromodulation: a cadaveric studyInternational Urogynecology Journal, 2014
- Randomized prospective crossover study of interstim lead wire placement with curved versus straight styletNeurourology and Urodynamics, 2013
- Optimizing electrode implantation in sacral nerve stimulation—an anatomical cadaver study controlled by a laparoscopic cameraInternational Journal of Colorectal Disease, 2007
- Predicting implantation with a neuromodulator using two different test stimulation techniques: A prospective randomized study in urge incontinent womenNeurourology and Urodynamics, 2006
- Failure of Sacral Nerve Stimulation Due to Migration of Tined LeadJournal of Urology, 2006
- Modified techniques of S3 foramen localization and lead implantation in S3 neuromodulationUrology, 2001
- Functional evaluation of sacralnerve root integrityUrology, 1990