Accuracy of heart rate variability estimation by photoplethysmography using an smartphone: Processing optimization and fiducial point selection
Open Access
- 1 August 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 2015 (1094687X) , 5700-5703
Abstract
This work compares several fiducial points to detect the arrival of a new pulse in a photoplethysmographic signal using the built-in camera of smartphones or a photoplethysmograph. Also, an optimization process for the signal preprocessing stage has been done. Finally we characterize the error produced when we use the best cutoff frequencies and fiducial point for smartphones and photopletysmograph and compare if the error of smartphones can be reasonably be explained by variations in pulse transit time. The results have revealed that the peak of the first derivative and the minimum of the second derivative of the pulse wave have the lowest error. Moreover, for these points, high pass filtering the signal between 0.1 to 0.8 Hz and low pass around 2.7 Hz or 3.5 Hz are the best cutoff frequencies. Finally, the error in smartphones is slightly higher than in a photoplethysmograph.Keywords
This publication has 9 references indexed in Scilit:
- iPhone 4s Photoplethysmography: Which Light Color Yields the Most Accurate Heart Rate and Normalized Pulse Volume Using the iPhysioMeter Application in the Presence of Motion Artifact?PLOS ONE, 2014
- Differences in QRS Locations due to ECG Lead: Relationship with Breathing2014
- Smartphone-enabled pulse rate variability: An alternative methodology for the collection of heart rate variability in psychophysiological researchInternational Journal of Psychophysiology, 2013
- Statistical analysis of heart rate and heart rate variability monitoring through the use of smart phone cameras2012
- Physiological Parameter Monitoring from Optical Recordings With a Mobile PhoneIEEE Transactions on Biomedical Engineering, 2012
- Errors in the Estimation of Approximate Entropy and Other Recurrence-Plot-Derived Indices Due to the Finite Resolution of RR Time SeriesIEEE Transactions on Biomedical Engineering, 2009
- Bias and uncertainty in heart rate variability spectral indices due to the finite ECG sampling frequencyPhysiological Measurement, 2004
- Sampling frequency of the electrocardiogram for spectral analysis of the heart rate variabilityIEEE Transactions on Biomedical Engineering, 1990
- A Real-Time QRS Detection AlgorithmIEEE Transactions on Biomedical Engineering, 1985