A custom-built PET phantom design for quantitative imaging of printed distributions
Authors:
- P J MarkiewiczPMP J Markiewicz
- G I AngelisGAG I Angelis
- F KotasidisFKF Kotasidis
- M GreenMGM Green
- W R LionheartWLW R Lionheart
- A J ReaderARA J Reader
- J C MatthewsJMJ C Matthews
- 7 October 2011
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (21) , N247-N261
Abstract
This note presents a practical approach to a custom-made design of PET phantoms enabling the use of digital radioactive distributions with high quantitative accuracy and spatial resolution. The phantom design allows planar sources of any radioactivity distribution to be imaged in transaxial and axial (sagittal or coronal) planes. Although the design presented here is specially adapted to the high-resolution research tomograph (HRRT), the presented methods can be adapted to almost any PET scanner. Although the presented phantom design has many advantages, a number of practical issues had to be overcome such as positioning of the printed source, calibration, uniformity and reproducibility of printing. A well counter (WC) was used in the calibration procedure to find the nonlinear relationship between digital voxel intensities and the actual measured radioactive concentrations. Repeated printing together with WC measurements and computed radiography (CR) using phosphor imaging plates (IP) were used to evaluate the reproducibility and uniformity of such printing. Results show satisfactory printing uniformity and reproducibility; however, calibration is dependent on the printing mode and the physical state of the cartridge. As a demonstration of the utility of using printed phantoms, the image resolution and quantitative accuracy of reconstructed HRRT images are assessed. There is very good quantitative agreement in the calibration procedure between HRRT, CR and WC measurements. However, the high resolution of CR and its quantitative accuracy supported by WC measurements made it possible to show the degraded resolution of HRRT brain images caused by the partial-volume effect and the limits of iterative image reconstruction.Keywords
This publication has 9 references indexed in Scilit:
- Image plate response for conditions relevant to laser–plasma interaction experimentsMeasurement Science and Technology, 2008
- Production of radioactive quality assurance phantoms using a standard inkjet printerPhysics in Medicine & Biology, 2007
- A new improved version of the realistic digital brain phantomNeuroImage, 2006
- OSEM-3D Reconstruction Strategies for the ECAT HRRT2005
- Printed sources for positron emission tomography (PET)IEEE Transactions on Nuclear Science, 2005
- Calibration of a Radioactive Ink-Based Stack Phantom and Its Applications in Nuclear MedicineCancer Biotherapy and Radiopharmaceuticals, 2003
- Experimental Performance Assessment of SPM for SPECT Neuroactivation Studies Using a Subresolution Sandwich Phantom DesignNeuroImage, 2002
- A novel phantom design for emission tomography enabling scatter- and attenuation-”free" single-photon emission tomography imagingEuropean Journal of Nuclear Medicine and Molecular Imaging, 2000
- Design and construction of a realistic digital brain phantomIEEE Transactions on Medical Imaging, 1998