CT Colonography: Protocol Optimization with Multi–Detector Row CT—Study in an Anthropomorphic Colon Phantom
Authors:
- Johannes WesslingJWJohannes Wessling
- Roman FischbachRFRoman Fischbach
- Norbert MeierNMNorbert Meier
- Thomas AllkemperTAThomas Allkemper
- Jutta KlusmeierJKJutta Klusmeier
- Karl LudwigKLKarl Ludwig
- Walter HeindelWHWalter Heindel
- 1 September 2003
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 228 (3) , 753-759
Abstract
To determine optimal detector collimation, section thickness, and tube current for multi-detector row computed tomography (CT) colonography. An anthropomorphic colon phantom with simulated polyps of varying size (2, 6, 8, 10, and 12 mm) was examined by using multi-detector row CT with varying combinations of detector collimation (4 x 1.0 mm and 4 x 2.5 mm), dose per section (10, 20, 40, 60, 80, 100, and 140 mAs), and section thickness/reconstruction interval (1.25/0.6, 2.0/1.0, 3.0/1.0, and 5.0/2.0 mm). Polyp depiction, longitudinal polyp distortion, and presence of rippling artifacts were assessed on reformatted three-dimensional endoluminal images by three reviewers. Longitudinal distortion and rippling artifacts increased with increasing section thickness and use of broader detector collimation. Polyps 8 mm or larger were depicted with any combination of section thickness, detector collimation, and tube current. Depiction of polyps 6 mm or smaller depended on the detector collimation/reconstructed section thickness and was rated optimal for the 4 x 1.0-mm detector collimation with a section thickness of 1.25 mm. This was also observed for low-dose protocols. Polyps 6 mm or smaller that were not detected with 3-mm section thickness and 4 x 2.5-mm detector collimation were detected with 1.25-mm section thickness and 10 mAs. A narrow detector collimation with thin-section imaging (4 x 1.0-mm detector collimation, 1.25-mm section thickness) is a prerequisite for low-dose (10-mAs) multi-detector row CT colonography.Keywords
This publication has 23 references indexed in Scilit:
- Colorectal polyps: Detection with multi-slice CT colonographyRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2001
- CT Colonography: Single- versus Multi–Detector Row ImagingRadiology, 2001
- Colonic Masses: Detection with MR ColonographyRadiology, 2000
- A Comparison of Virtual and Conventional Colonoscopy for the Detection of Colorectal PolypsNew England Journal of Medicine, 1999
- Subsecond multi-slice computed tomography: basics and applicationsEuropean Journal of Radiology, 1999
- A PC program for estimating organ dose and effective dose values in computed tomographyEuropean Radiology, 1999
- Detection of Colonic Polyps in a Phantom Model: Implications for Virtual Colonoscopy Data AcquisitionJournal of Computer Assisted Tomography, 1998
- Reducing data size and radiation dose for CT colonography.American Journal of Roentgenology, 1997
- Small simulated polyps in pig colon: sensitivity of CT virtual colography.Radiology, 1997
- Detection of colorectal polyps by computed tomographic colography: Feasibility of a novel techniqueGastroenterology, 1996