Abstract
Ibuprofen membrane electrodes based on different plasticizers: diisobutyl phthalate (DIBP), o-nitrophenyloctyl ether (o-NPOE), dioctyl sebacate (DOS) and tetraoctylammonium 2-(4-isobutylphenyl)propionate were prepared. All electrodes show: a near Nernstian slope of characteristic (58.3–60.9 mV decade−1) in the measurement range (10−4–10−1 mol L−1), limit of detection (5.0×10−5 mol L−1), really long lifetime (12 months), dependence of the electrode potential on pH (5.5–9.0), reproducibility of potential (0.6–1.2 mV) and selectivity coefficients in relation to some organic and inorganic anions. The electrodes were applied for the determination of ibuprofen in tablets by the calibration curve method and the standard addition method.Keywords
This publication has 30 references indexed in Scilit:
- Properties of naproxen ion-selective electrodesOpen Chemistry, 2008
- Performance evaluation criteria for preparation and measurement of macro- and microfabricated ion-selective electrodes (IUPAC Technical Report)Pure and Applied Chemistry, 2008
- Ion-selective electrodes in medicinal drug determinationRussian Chemical Reviews, 2007
- A new potentiometric ibuprofenate ion sensor immobilized in a graphite matrix for determination of ibuprofen in tabletsMicrochemical Journal, 2006
- Synthesis, pharmacological activity and hydrolytic behavior of glyceride prodrugs of ibuprofenEuropean Journal of Medicinal Chemistry, 2005
- Simultaneous spectrophotometric determination of pseudoephedrine hydrochloride and ibuprofen in a pharmaceutical preparation using ratio spectra derivative spectrophotometry and multivariate calibration techniquesJournal of Pharmaceutical and Biomedical Analysis, 2004
- The non-aqueous titrimetric assay of the selected anti-inflammatory agents using tetra-n-butylammonium hydroxide as titrantJournal of Pharmaceutical and Biomedical Analysis, 1999
- Direct potentiometry and potentiotitrimetry of warfarin and ibuprofen in pharmaceutical preparations using PVC ferroin-based membrane sensorsMicrochimica Acta, 1998
- Ion-Selective Membrane Electrodes in Pharmaceutical AnalysisCritical Reviews in Analytical Chemistry, 1997
- Spectrophotometric Determination of Ibuprofen via its Copper (II) ComplexAnalytical Letters, 1994