Abstract
This letter presents a capacitively coupled chopping instrumentation amplifier (CCIA) with an orthogonal timing mismatch averaging (OTMA) technique for biosignal sensing. The OTMA employs additional pairs of choppers in input and feedback networks to mitigate all the passive component mismatches through four orthogonal phases. The OTMA substantially increases the common-mode rejection ratio (CMRR) without the need for complex correction loops or extra pathways. Designed in 130-nm CMOS technology, the proposed CCIA with the OTMA draws 3.5 μA from a 1.8-V supply and exhibits 0.63-μVrms input-referred noise over 0.5–200 Hz. It achieves >130-dB CMRR with a compact chip area.

This publication has 14 references indexed in Scilit: