Duplicate measures of hemoglobin mass within an hour: feasibility, reliability, and comparison of three devices in supine position
- 24 January 2024
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 84 (1) , 1-10
Abstract
Duplicate measure of hemoglobin mass by carbon monoxide (CO)-rebreathing is a logistical challenge as recommendations prompt several hours between measures to minimize CO-accumulation. This study investigated the feasibility and reliability of performing duplicate CO-rebreathing procedures immediately following one another. Additionally, it was evaluated whether the obtained hemoglobin mass from three different CO-rebreathing devices is comparable. Fifty-five healthy participants (22 males, 23 females) performed 222 duplicate CO-rebreathing procedures in total. Additionally, in a randomized cross-over design 10 participants completed three experimental trials, each including three CO-rebreathing procedures, with the first and second separated by 24 h and the second and third separated by 5-10 min. Each trial was separated by >48 h and conducted using either a glass-spirometer, a semi-automated electromechanical device, or a standard three-way plastic valve designed for pulmonary measurements. Hemoglobin mass was 3 ± 22 g lower (p < 0.05) at the second measure when performed immediately after the first with a typical error of 1.1%. Carboxyhemoglobin levels reached 10.9 ± 1.3%. In the randomized trial, hemoglobin mass was similar between the glass-spirometer and three-way valve, but ∼6% (∼50 g) higher for the semi-automated device. Notably, differences in hemoglobin mass were up to ∼13% (∼100 g) when device-specific recommendations for correction of CO loss to myoglobin and exhalation was followed. In conclusion, it is feasible and reliable to perform two immediate CO-rebreathing procedures. Hemoglobin mass is comparable between the glass-spirometer and the three-way plastic valve, but higher for the semi-automated device. The differences are amplified if the device-specific recommendations of CO-loss corrections are followed.Keywords
This publication has 42 references indexed in Scilit:
- Comparability of haemoglobin mass measured with different carbon monoxide-based rebreathing procedures and calculationsScandinavian Journal of Clinical and Laboratory Investigation, 2010
- Carbon monoxide uptake kinetics of arterial, venous and capillary blood during CO rebreathingExperimental Physiology, 2010
- Stability of hemoglobin mass over 100 days in active menJournal of Applied Physiology, 2008
- Loss of CO from the intravascular bed and its impact on the optimised CO-rebreathing methodEuropean Journal of Applied Physiology, 2007
- Time and Sample Site Dependency of the Optimized CO-Rebreathing MethodMedicine & Science in Sports & Exercise, 2006
- Errors of measurement for blood volume parameters: a meta-analysisJournal of Applied Physiology, 2005
- The optimised CO-rebreathing method: a new tool to determine total haemoglobin mass routinelyEuropean Journal of Applied Physiology, 2005
- The Effect of Carbon Monoxide on HumansAnnual Review of Pharmacology, 1975
- Determination of Circulating Red Cell Volume by Radioactive ChromiumScience, 1950
- CLINICAL STUDIES OF THE BLOOD VOLUME. I. CLINICAL APPLICATION OF A METHOD EMPLOYING THE AZO DYE “EVANS BLUE” AND THE SPECTROPHOTOMETERJournal of Clinical Investigation, 1937