Dosing accuracy of two insulin pre-filled pens
- 8 April 2008
- journal article
- research article
- Published by Informa Healthcare in Current Medical Research and Opinion
- Vol. 24 (5) , 1429-1434
Abstract
Objective: This study was designed to determine the dose accuracy of two commonly available insulin pre-filled pens for use in diabetes – FlexPen (Novo Nordisk A/S, Bagsværd, Denmark) (FP) and SoloStar (Sanofi-Aventis, Paris, France) (SS).Research design and methods: Dosing accuracy was tested at 5 U, 10 U and 30 U doses – three previously unused pens of both pre-filled pens were used for each dose. At the 5 U dose each pen was tested 42 times, at the 10 U dose each pen was tested 25 times and at the 30 U dose each pen was tested 9 times. The pre-filled pens were used strictly according to manufacturers’ instructions and measurements were made on a sensitive balance and corrected for the specific density of the insulin formulations used. Specified limits were based on ISO standards (±1 U for the 5 U and 10 U doses and ±1.5 U for the 30 U dose).Results: FP was more accurate for injecting 5 U, 10 U and 30 U doses (absolute mean 4.95 ± 0.19 U, 9.61 ± 0.27 U and 29.70 ± 0.34 U, respectively) than SS (...Keywords
This publication has 6 references indexed in Scilit:
- Insulin Pen Injection Devices for Management of Patients with Type 2 Diabetes: Considerations Based on an Endocrinologist’s Practical Experience in the United StatesEndocrine Practice, 2007
- Dose accuracy and injection force dynamics of a novel disposable insulin penExpert Opinion on Drug Delivery, 2007
- A multicenter, randomized, open-label, comparative, two-period crossover trial of preference, efficacy, and safety profiles of a prefilled, disposable pen and conventional vial/syringe for insulin injection in patients with type 1 or 2 diabetes mellitusClinical Therapeutics, 2003
- Accuracy of pen injectors versus insulin syringes in children with type 1 diabetes.Diabetes Care, 1999
- Safety, efficacy, acceptability of a pre-filled insulin pen in diabetic patients over 60 years oldDiabetes Research and Clinical Practice, 1995
- Dosage Accuracy of Self-mixed vs Premixed InsulinArchives of internal medicine (1960), 1991