Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes
- 3 September 2019
- journal article
- research article
- Published by Springer Nature in Graefe's Archive for Clinical and Experimental Ophthalmology
- Vol. 257 (11) , 2437-2447
Abstract
Purpose To evaluate the intrasession repeatability of anterior corneal topographic and aberrometric measurements provided by a color-LED topographer as well as their interchangeability with those provided by a Scheimpflug-based system in healthy eyes. Methods Thirty-five healthy eyes of 35 patients (age, 16–66 years) were enrolled. A complete eye examination was performed in all cases including a complete corneal analysis with the Scheimpflug-based system Pentacam (Oculus Optikgeräte) (one measurement) and the Cassini system (i-Optics) (three consecutive measurements). Intrasession repeatability of the Cassini measurements was assessed with the within-subject standard deviation (Sw) and the intraclass correlation coefficient (ICC). The Bland-Altman analysis was used to evaluate the agreement between both devices. Results Mean Sw for keratometric readings was 0.02 mm (ICC ≥ 0.992), ranging between 0.16 and 0.05 D (ICC 0.930–0.978) for anterior and total astigmatic measurements. Mean Sw for asphericity and corneal diameter were 0.06 (ICC 0.926) and 0.03 mm (IC 0.997), respectively. Aberrometric parameters showed ICCs ≥ 0.816, except for Z42 (ICC 0.741) and Z44 (ICC 0.544). When comparing devices, statistically significant differences were found for most of topographic and aberrometric data (p ≤ 0.044). Likewise, ranges of agreement between devices were clinically relevant (keratometry > 0.06 mm; total astigmatic components > 0.69 D; asphericity 0.35; second-, third-, and fourth-order Zernike terms, more than 0.20, 0.13, and 0.01 μm, respectively). Conclusions Consistent anterior corneal topographic, total corneal astigmatic, and aberrometric measurements are obtained with color-LED topography in healthy eyes, which are not interchangeable with those provided by the Scheimpflug-based topography.Keywords
Funding Information
- Ministerio de Economía, Industria y Competitividad, Gobierno de España (RYC-2016-20471)
This publication has 33 references indexed in Scilit:
- Repeatability of aberrometric measurements in normal and keratoconus eyes using a new Scheimpflug–Placido topographerJournal of Cataract & Refractive Surgery, 2014
- Technologies for Anatomical and Geometric Characterization of the Corneal Structure and Anterior Segment: A ReviewSeminars in Ophthalmology, 2013
- Forme Fruste Keratoconus Imaging and Validation via Novel Multi-Spot Reflection TopographyCase Reports in Ophthalmology, 2013
- Correlation of the Corneal Toricity Between Anterior and Posterior Corneal Surfaces in the Normal Human EyeCornea, 2013
- Intrasubject repeatability of corneal morphology measurements obtained with a new Scheimpflug photography–based systemJournal of Cataract & Refractive Surgery, 2012
- Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topographyJournal of Cataract & Refractive Surgery, 2011
- Repeatability of corneal power and wavefront aberration measurements with a dual-Scheimpflug Placido corneal topographerJournal of Cataract & Refractive Surgery, 2010
- Repeatability of corneal first-surface wavefront aberrations measured with Pentacam corneal topographyJournal of Cataract & Refractive Surgery, 2008
- Intrasession and intersession repeatability of the Pentacam system on posterior corneal assessment in the normal human eyeJournal of Cataract & Refractive Surgery, 2007
- A new instrument to measure the shape of the cornea based on pseudorandom color codingIEEE Transactions on Instrumentation and Measurement, 1997