Rotary powered device for bone marrow aspiration and biopsy yields excellent specimens quickly and efficiently
Authors:
- 19 April 2010
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 63 (6) , 562-565
Abstract
Recently, a new FDA-cleared battery powered bone marrow biopsy system was developed to allow operators access to the bone marrow space quickly and efficiently. A pre-clinical evaluation of the device (OnControl, Vidacare Corporation, San Antonio, TX, USA) on anesthetized pigs was conducted, in addition to a clinical evaluation in hematology clinic patients requiring a bone marrow biopsy. Twenty-six samples were collected from the swine model. No cellular artifact or thermal damage was reported in any of the samples obtained. For the clinical evaluation of the device, 16 patients were recruited. Mean time from needle contact with skin to needle removal was 38.5 +/− 13.94 seconds. No complications were reported. In this study, the manual and powered samples were equivalent in specimen quality. In the patients evaluated, the device was safe, easy to use and the mean procedural time was significantly faster than previously reported with a manual technique.Keywords
This publication has 5 references indexed in Scilit:
- Assessment of bone biopsy needles for sample size, specimen quality and ease of useSkeletal Radiology, 2005
- Bone marrow punctures and painAcute Pain, 2004
- Bone marrow biopsy morbidity and mortalityBritish Journal of Haematology, 2003
- Audit of bone marrow trephines.Journal of Clinical Pathology, 1992
- Value of long‐core biopsy in the detection of discrete bone marrow lesionsHistopathology, 1988