SIMPLE METHOD OF SIZE-SPECIFIC DOSE ESTIMATES CALCULATION FROM PATIENT WEIGHT ON COMPUTED TOMOGRAPHY
- 28 July 2017
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 178 (2) , 208-212
Abstract
The purpose of this study is to develop a new and simple methodology for calculating mean size-specific dose estimates (SSDE) over the entire scan range (mSSDE) from weight and volume CT dose index (CTDIvol). We retrospectively analyzed data from a dose index registry. Scan areas were divided into two regions: chest and abdomen–pelvis. The original mSSDE was calculated by a commercially available software. The conversion formulas for mSSDE were estimated from weight and CTDIvol (SSDEweight) in each region. SSDEweight were compared with the original mSSDE using Bland–Altman analysis. Root mean square differences were 1.4 mGy for chest and 1.5 mGy for abdomen–pelvis. Our method using formulae can calculate SSDEweight using weight and CTDIvol without a dedicated software, and can be used to calculate DRL to optimize CT exposure doses.Keywords
This publication has 14 references indexed in Scilit:
- Size-Specific Dose Estimates for Evaluation of Individual Patient Dose in CT Protocol for Renal ColicAmerican Journal of Roentgenology, 2015
- Size-specific Dose Estimates for Chest, Abdominal, and Pelvic CT: Effect of Intrapatient Variability in Water-equivalent DiameterRadiology, 2015
- Simplifying Size-Specific Radiation Dose Estimates in Pediatric CTAmerican Journal of Roentgenology, 2015
- Local diagnostic reference level based on size-specific dose estimates: Assessment of pediatric abdominal/pelvic computed tomography at a Japanese national children’s hospitalPediatric Radiology, 2014
- Size-specific dose estimate (SSDE) provides a simple method to calculate organ dose for pediatric CT examinationsMedical Physics, 2014
- Estimating Patient Dose from X-ray Tube Output Metrics: Automated Measurement of Patient Size from CT Images Enables Large-scale Size-specific Dose EstimatesRadiology, 2014
- Investigation of the freely available easy-to-use software ‘EZR’ for medical statisticsBone Marrow Transplantation, 2012
- Size-specific Dose Estimates for Adult Patients at CT of the TorsoRadiology, 2012
- Statistical methods for assessing agreement between two methods of clinical measurementInternational Journal of Nursing Studies, 2010
- A method for describing the doses delivered by transmission x-ray computed tomographyMedical Physics, 1981