Radiation Dosimetry Descriptors Applied to Four C T Scanners in New Zealand
Authors:
- 1 May 1984
- journal article
- Published by Wiley in Australasian Radiology
- Vol. 28 (2) , 161-170
Abstract
Doses to patients have been estimated for the four New Zealand CT Scanners, in terms of recently proposed dose descriptors. These descrip‐tors were computed tomography dose index (CTDI), multiple scan average dose (MSAD), “equivalent rectangular width (w), a figure of merit, and integral dose. X‐ray beam dose profiles, half‐value‐layer and peak kilovoltage were also determined. Dosimetry was by thermoluminescent dosimeters, pencil ionization chamber, and x‐ray films, in cylindrical solid perspex phantoms. The four scanners were all found to give comparable doses to patients.Keywords
This publication has 7 references indexed in Scilit:
- Dose descriptors for computed tomographyMedical Physics, 1982
- Radiation dosimetry survey of computed tomography systems from ten manufacturersBritish Journal of Radiology, 1982
- An evaluation of CT systems from ten manufacturersBritish Journal of Radiology, 1981
- A method for describing the doses delivered by transmission x-ray computed tomographyMedical Physics, 1981
- Dosimetry of C.T. ScannersAustralasian Radiology, 1980
- Radiation Dosimetry In Computed Tomography (CT)1977
- Checking diagnostic X-ray beam qualityBritish Journal of Radiology, 1968