Pediatric Cardiovascular CT Angiography: Radiation Dose Reduction Using Automatic Anatomic Tube Current Modulation

Authors:

Abstract
OBJECTIVE. The purpose of this study was to assess the effect of weight-based scanning protocols and automatic tube current modulation on the tube current–time product and image quality at pediatric cardiovascular 64-MDCT angiography.MATERIALS AND METHODS. Our pediatric cardiovascular 64-MDCT protocols use a weight-based algorithm to determine nominal tube voltage settings with 80, 100, and 120 kV. Automatic tube current modulation was used for each case. The mAs, volume CT dose index (CTDIvol), and dose–length product (DLP) values were recorded and the effective dose calculated. On the basis of the selected nominal tube current, the dose values that would have been delivered without tube current modulation were also calculated. Scans were compared with 16-MDCT using 120 kVp and 120 mAs. Two radiologists independently rated image quality on a 5-point scale. Image noise was objectively measured within four different regions of interest. Findings at CT were clinically correlated with results of cardiac sono...