Evaluation of a High-Sensitivity Organ-Targeted PET Camera
Authors:
Open Access
Abstract
The aim of this study is to evaluate the performance of the Radialis organ-targeted positron emission tomography (PET) Camera with standardized tests and through assessment of clinical-imaging results. Sensitivity, count-rate performance, and spatial resolution were evaluated according to the National Electrical Manufacturers Association (NEMA) NU-4 standards, with necessary modifications to accommodate the planar detector design. The detectability of small objects was shown with micro hotspot phantom images. The clinical performance of the camera was also demonstrated through breast cancer images acquired with varying injected doses of 2-[fluorine-18]-fluoro-2-deoxy-D-glucose (18F-FDG) and qualitatively compared with sample digital full-field mammography, magnetic resonance imaging (MRI), and whole-body (WB) PET images. Micro hotspot phantom sources were visualized down to 1.35 mm-diameter rods. Spatial resolution was calculated to be 2.3 ± 0.1 mm for the in-plane resolution and 6.8 ± 0.1 mm for the cross-plane resolution using maximum likelihood expectation maximization (MLEM) reconstruction. The system peak noise equivalent count rate was 17.8 kcps at a 18F-FDG concentration of 10.5 kBq/mL. System scatter fraction was 24%. The overall efficiency at the peak noise equivalent count rate was 5400 cps/MBq. The maximum axial sensitivity achieved was 3.5%, with an average system sensitivity of 2.4%. Selected results from clinical trials demonstrate capability of imaging lesions at the chest wall and identifying false-negative X-ray findings and false-positive MRI findings, even at up to a 10-fold dose reduction in comparison with standard 18F-FDG doses (i.e., at 37 MBq or 1 mCi). The evaluation of the organ-targeted Radialis PET Camera indicates that it is a promising technology for high-image-quality, low-dose PET imaging. High-efficiency radiotracer detection also opens an opportunity to reduce administered doses of radiopharmaceuticals and, therefore, patient exposure to radiation.Keywords
Funding Information
- Canadian Cancer Society (CCS)
- Ontario Research Fund—Research Excellence Program (ORF-RE)
- Natural Sciences and Engineering Research Council of Canada (NSERC)
This publication has 49 references indexed in Scilit:
- Diagnostic Performance of Dedicated Positron Emission Mammography Using Fluorine-18-Fluorodeoxyglucose in Women With Suspicious Breast Lesions: A Meta-analysisClinical Breast Cancer, 2014
- A Molecular Approach to Breast ImagingJournal of Nuclear Medicine, 2014
- Nuclear imaging of the breast: Translating achievements in instrumentation into clinical useMedical Physics, 2013
- Design and evaluation of the MAMMI dedicated breast PETMedical Physics, 2012
- Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR ScannerJournal of Nuclear Medicine, 2011
- Comparison of radiation exposure and associated radiation‐induced cancer risks from mammography and molecular imaging of the breasta)Medical Physics, 2010
- Radiation Doses and Cancer Risks from Breast Imaging StudiesRadiology, 2010
- Clinical Imaging Characteristics of the Positron Emission Mammography Camera: PEM Flex Solo IIJournal of Nuclear Medicine, 2009
- Diagnostic Performance of Digital versus Film Mammography for Breast-Cancer ScreeningNew England Journal of Medicine, 2005
- Bayesian image reconstruction for emission tomography based on median root priorEuropean Journal of Nuclear Medicine and Molecular Imaging, 1997