Safety, Efficacy, and Performance of Implanted Recycled Cardiac Rhythm Management (CRM) Devices in Underprivileged Patients
- 21 March 2011
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 34 (6) , 653-658
Abstract
Patients in underdeveloped nations have limited access to life-saving medical technology including cardiac rhythm management (CRM) devices. We evaluated alternative means to provide such technology to this patient population while assessing the safety and efficacy of such a practice. Patients in the United States with clinical indications for extraction of CRM devices were consented. Antemortem CRM devices were cleaned and sterilized following a protocol established at our institution. Surveillance in vitro cultures were performed for quality assurance. The functional status of pulse generators was tested with a pacing system analyzer to confirm at least 70% battery life. Most generators were transported, in person, to an implanting institution in Nicaragua. Recipients with a Class I indication for CRM implantation, and meeting economical criteria set forth, were consented for implantation of a recycled device. Between 2003 and 2009, implantation was performed in 17 patients with an average age of 42.1 ± 20.3 years. Of the 17 patients, nine were male and eight were female. Mean follow-up was 68 ± 38 months. Device evaluation occurred prior to discharge, 4 weeks post implantation, and every 6 months thereafter. There were three deaths during the follow-up period secondary to myocardial infarction, stroke, and heart failure. Hematoma formation occurred in one patient. No infections, early battery depletion, or device malfunction were identified during follow-up. Our case series is the longest follow-up of recipients of recycled antemortem CRM devices. Our findings support the feasibility and safety of this alternative acquisition of life-saving technology.Keywords
This publication has 9 references indexed in Scilit:
- Abstract P96: Pacemaker Reutilization is a Safe and Effective Means of Delivering Electrophysiological Healthcare in Third World Countries a Meta-AnalysisCirculation: Cardiovascular Quality and Outcomes, 2011
- Reuse of pacemakers and defibrillators in developing countries: Logistical, legal, and ethical barriers and solutionsHeart Rhythm, 2010
- Pacemaker ReuseCirculation, 2010
- Safety and Efficacy of Pacemaker Reuse in Underdeveloped NationsJACC, 2009
- Heartbeat International: Making “poor” hearts beat betterHeart Rhythm, 2009
- ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm AbnormalitiesCirculation, 2008
- Projections of Global Mortality and Burden of Disease from 2002 to 2030PLOS Medicine, 2006
- ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices—Summary ArticleJACC, 2002
- Re-used pacemakers—as safe as new? A retrospective case-control studyEuropean Heart Journal, 1998