Comparison of accuracy and axial length acquisition success rate by three types of swept-source OCT-based biometers
Open Access
- 16 December 2024
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 51 (4) , 287-293
Abstract
Purpose: To compare three biometers equipped with swept-source optical coherence tomography (OCT), including ARGOS (OCTB1), IOLMaster 700 (OCTB2), and Anterion (AS-OCTB). The primary aim was to assess the axial length (AL) acquisition success rates, and secondary aims included comparing parameters obtained from the three devices and evaluating postoperative refractive prediction errors. Setting: Chukyo Eye Clinic, Nagoya, Japan. Design: Retrospective observational study. Methods: The study included patients undergoing preoperative assessments for cataract surgery using the three biometers. The AL acquisition success rates and parameters obtained from the three devices were compared and the postoperative refractive prediction errors (defined as the difference between subjective refractive error and predicted refractive error) were assessed at one month after surgery. Results: The study included 361 eyes from 361 patients. The AL acquisition success rates for all patients were 100.0% (n=361) (OCTB1), 98.6% (n=356) (OCTB2), and 96.7% (AS-OCTB) (P < .0001), and those for nuclear sclerosis of grade 4 or higher were 100.0%, 87.8%, and 87.8%, respectively (P = .007). The AL of mature cataract was measurable only with OCTB1 using the enhanced retina visualization mode. Although there were significant differences in parameters obtained on the three devices, there were no significant differences in postoperative refractive prediction errors. Conclusions: OCTB1 demonstrated the highest AL acquisition success rate and was able to measure AL even in eyes with dense or mature cataract. AS-OCTB using a wavelength of 1300 nm showed lower AL acquisition success rates compared to the other two devices probably due to reduced light transmission through the vitreous body.Keywords
This publication has 20 references indexed in Scilit:
- Accuracy of Intraocular Lens Calculation FormulasOphthalmology, 2018
- Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography–based biometer versus partial coherence interferometry and optical low-coherence reflectometryJournal of Cataract & Refractive Surgery, 2016
- Biometry measurements using a new large-coherence–length swept-source optical coherence tomographerJournal of Cataract & Refractive Surgery, 2016
- Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500British Journal of Ophthalmology, 2015
- Clinical comparison of a new swept-source optical coherence tomography–based optical biometer and a time-domain optical coherence tomography–based optical biometerJournal of Cataract & Refractive Surgery, 2015
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 2008
- Immersion A-scan compared with partial coherence interferometryJournal of Cataract & Refractive Surgery, 2002
- Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to HaigisGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- Partial coherence interferometry: a novel approach to biometry in cataract surgeryAmerican Journal of Ophthalmology, 1998
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992