HeartMate III: Pump Design for a Centrifugal LVAD with a Magnetically Levitated Rotor

Authors:
  • DGDavid B. Gernes
    David B. Gernes
  • HLHoward M. Loree
    Howard M. Loree
  • JRJ. Scott Richardson
    J. Scott Richardson
  • VPVictor L. Poirier
    Victor L. Poirier
  • NBNatale Barletta
    Natale Barletta
  • AFAndreas Fleischli
    Andreas Fleischli
  • GFGiampiero Foiera
    Giampiero Foiera
  • TGThomas M. Gempp
    Thomas M. Gempp
  • RSReto Schoeb
    Reto Schoeb
  • KLKenneth N. Litwak
    Kenneth N. Litwak
  • TATakehide Akimoto
    Takehide Akimoto

Abstract
A long-term, compact left ventricular assist device (LVAD), the HeartMate III, has been designed and fabricated, featuring a centrifugal pump with a magnetically levitated rotor. The pump has been optimized by in vitro testing to achieve a design point of 7 L/min against 135 mm Hg at high hydrodynamic efficiency (30%) and to be capable of up to 10 L/min under such a load. Furthermore, the pump has demonstrated no mechanical failures, low hemolysis (4–10 mg/dl plasma free Hb), and low thrombogenicity during six (40, 27, 59, 42, 27, and 49-day) in vivo bovine studies.