Compensation of temperature effects in quartz crystal microbalance measurements
Authors:
- 21 January 2002
- journal article
- conference paper
- Published by AIP Publishing in Applied Physics Letters
- Vol. 80 (3) , 521-523
Abstract
Quartz crystal microbalance (QCM) is a very powerful method for in situ monitoring of thin film growth processes. However, especially at high temperatures, QCM is very sensitive to already minor temperature changes during the measurement. Here, a method for compensating the temperature effects on the QCM is introduced. In this method, the baseline drift during the growth is calculated using a function obtained by fitting the signal measured before and after the growth. The present method is compared with the earlier published reference crystal method where two closely spaced quartz crystals are used, one being protected against film growth and thus serving as a reference crystal for compensating the temperature effects on the measurement crystal. Both methods were tested with and atomic layer deposition processes and the temperature effect was successfully compensated. The differences between these methods are discussed.
Keywords
This publication has 12 references indexed in Scilit:
- In Situ Quartz Crystal Microbalance and Quadrupole Mass Spectrometry Studies of Atomic Layer Deposition of Aluminum Oxide from Trimethylaluminum and WaterLangmuir, 2001
- Atomic layer deposition of titanium dioxide from TiCl4 and H2O: investigation of growth mechanismApplied Surface Science, 2001
- Study of atomic layer epitaxy of zinc oxide by in-situ quartz crystal microgravimetryApplied Surface Science, 2000
- Perfectly Conformal TiN and Al2O3 Films Deposited by Atomic Layer DepositionChemical Vapor Deposition, 1999
- Synthesis of oxide thin films and overlayers by atomic layer epitaxy for advanced applicationsMaterials Science and Engineering: B, 1996
- Development and testing of a high temperature quartz crystal microbalanceVacuum, 1995
- Thickness-shear-mode acoustic wave sensors in the liquid phase. A reviewThe Analyst, 1991
- In Situ Interfacial Mass Detection with Piezoelectric TransducersScience, 1990
- A Quartz-Crystal Microbalance for Monitoring Vacuum Depositions at Elevated Temperatures1966
- Verwendung von Schwingquarzen zur W gung d nner Schichten und zur Mikrow gungThe European Physical Journal A, 1959