Artificial Urinary Sphincter: Long-Term Results and Patient Satisfaction
Open Access
- 26 March 2012
- journal article
- review article
- Published by Hindawi Limited in Advances in Urology
- Vol. 2012, 1-4
Abstract
The published evidence concerning the safety, efficacy, and patient satisfaction for implantation of the current model of the artificial urinary sphincter (AS 800) in men with post prostatectomy urinary incontinence was the objective of this review. A Pub Med English language literature search from 1995 to 2011 was performed. A majority of men who undergo AUS implantation for post prostatectomy urinary incontinence achieve satisfactory results (0 to 1 pad per day). Infection rates range from 0.46 to 7%, cuff erosion rates range from 3.8 to 10%, and urethral atrophy ranges from 9.6 to 11.4%. Kaplan-Meier 5 year projections for freedom from any reoperation were 50% for a small series and 79.4% for a larger series. Kaplan-Meier projections for freedom from mechanical failure were 79% at 5 years and 72% at 10 years. In another series 10 year projections for freedom from mechanical failure were 64%. Although the artificial urinary sphincter (AUS) is the gold standard for the treatment of this disorder, most men will continue to need at least one pad per day for protection, and they are subject to a significant chance of future AUS revision or replacement.Keywords
This publication has 29 references indexed in Scilit:
- Surgical treatment of stress incontinence in menNeurourology and Urodynamics, 2009
- Long-Term Durability and Functional Outcomes Among Patients With Artificial Urinary Sphincters: A 10-Year Retrospective Review From the University of MichiganJournal of Urology, 2008
- 13 Years of Experience With Artificial Urinary Sphincter Implantation at Baylor College of MedicineJournal of Urology, 2007
- CHAPTER 1: THE MANAGEMENT OF ERECTILE DYSFUNCTION: AN AUA UPDATEJournal of Urology, 2005
- OUTCOMES FOLLOWING REVISIONS AND SECONDARY IMPLANTATION OF THE ARTIFICIAL URINARY SPHINCTERJournal of Urology, 2005
- Transcorporal Artificial Urinary Sphincter Cuff Placement in Cases Requiring Revision for Erosion and Urethral AtrophyJournal of Urology, 2002
- LONG-TERM CONTINENCE AND PATIENT SATISFACTION AFTER ARTIFICIAL SPHINCTER IMPLANTATION FOR URINARY INCONTINENCE AFTER PROSTATECTOMYJournal of Urology, 2001
- THE LONG-TERM OUTCOME OF ARTIFICIAL URINARY SPHINCTERSJournal of Urology, 2000
- Quality of Life and Continence Assessment of the Artificial Urinary Sphincter in Men With Minimum 3.5 Years of FollowupJournal of Urology, 1997
- Post-Prostatectomy incontinence and the Artificial Urinary Sphincter: A Long-Term Study of Patient Satisfaction and Criteria for SuccessJournal of Urology, 1996