A new mobile self-dispensing and administering system for 18 F-FDG: evaluation of operator dose reduction
Authors:
- 28 January 2020
- journal article
- research article
- Published by IOP Publishing in Journal of Radiological Protection
- Vol. 40 (1) , 243-252
Abstract
Purpose: Recently new mobile systems for dispensing positron emitters have been produced, designed to guarantee dispensing cycles in an aseptic environment. The aim of the present work was to assess the advantage of one of these systems in radiation protection of operators in clinical settings. Methods: Recently, in our centre the new self-dispensing system named KARL 100 by Tema Sinergie was adopted for 18 F-FDG radiopharmaceuticals. The system is associated with an automatic Rad-inject infuser. The system that was previously used was a fixed isolator NMC DSI (Tema Sinergie), equipped with a μ DDS-An activity fractioning system, together with a pneumatic post for the syringe delivery. The dosimetric evaluations on both systems were carried out through environmental measurements with an ionisation chamber and with the use of personal dosimeters. Results: The operations of preparation and administration of 18 F-FDG dose to the patient, with the use of Karl 100 + RadInject, involve exposures much lower than those obtained by the fixed isolator. The average body exposure of the technician was reduced by 31%, and for the physician by 77%. On the extremities, the equivalent dose to the hands of the technician was reduced by 78%, and for the physician by 96%. Also the additional dosimeters worn by the technician confirmed the estimated environmental assessments. Conclusions: The exposures of the working personnel were significantly reduced with the introduction of the new KARL 100 system.Keywords
This publication has 11 references indexed in Scilit:
- Strategies for assessment of doses to the tips of the fingers in nuclear medicineJournal of Radiological Protection, 2016
- Investigation of radiation protection of medical staff performing medical diagnostic examinations by using PET/CT techniqueJournal of Radiological Protection, 2015
- Trends in radiation protection of positron emission tomography/computed tomography imagingAnnals of the ICRP, 2014
- Validation of an automated dose-dispensing system for 18F-FDG administrations and associated reduction in operator extremity doseNuclear Medicine Communications, 2014
- Occupational Radiation Dosimetry Assessment Using an Automated Infusion Device for Positron-Emitting RadiotracersJournal of Nuclear Medicine Technology, 2012
- Validation of a new protocol for 18F-FDG infusion using an automatic combined dispenser and injector systemEuropean Journal of Nuclear Medicine and Molecular Imaging, 2012
- Hand exposure in diagnostic nuclear medicine with 18F- and 99mTc-labelled radiopharmaceuticals - Results of the ORAMED projectRadiation Measurements, 2011
- Extremity exposure in nuclear medicine: preliminary results of a European studyRadiation Protection Dosimetry, 2011
- The introduction of automated dispensing and injection during PET procedures: a step in the optimisation of extremity doses and whole-body doses of nuclear medicine staffRadiation Protection Dosimetry, 2010
- Personal dose monitoring in hospitals: global assessment, critical applications and future needsRadiation Protection Dosimetry, 2007