The use of monitoring end tidal carbon dioxide pressure (PetCO2) in neonatal intensive care was studied in 19 infants with respiratory disease. PetCO2 correlated poorly with arterial pCO2, the relation being principally determined by the severity of pulmonary disease. Monitoring end tidal CO2 cannot be recommended for neonates with pulmonary disease.Keywords
This publication has 4 references indexed in Scilit:
- End-tidal PCO2 greater than PaCO2Critical Care Medicine, 1984
- DEADSPACE AND THE SINGLE BREATH TEST FOR CARBON DIOXIDE DURING ANAESTHESIA AND ARTIFICIAL VENTILATIONBritish Journal of Anaesthesia, 1984
- Relationship between Arterial and Peak Expired Carbon Dioxide Pressure during Anesthesia and Factors Influencing the DifferenceAnesthesia & Analgesia, 1981
- Difference between end-tidal and arterial PCO2 in exerciseJournal of Applied Physiology, 1979