Precision and accuracy of stress measurement with a portable X-ray machine using an area detector
Open Access
- 1 February 2017
- journal article
- research article
- Published by International Union of Crystallography (IUCr) in Journal of Applied Crystallography
- Vol. 50 (1) , 131-144
Abstract
The use of portable X-ray stress analyzers, which utilize an area detector along with the newly adopted `cosα' or full-ring fitting method, has recently attracted increasing interest. In laboratory conditions, these measurements are fast, convenient and precise because they employ a single-exposure technique that does not require sample rotation. In addition, the effects of grain size and orientation can be evaluated from the Debye ring recorded on the area detector prior to data analysis. The accuracy of the measured stress, however, has been questioned because in most cases just a single reflection is analyzed and the sample-to-detector distances are relatively short. This article presents a comprehensive analysis of the uncertainty associated with a state-of-the-art commercial portable X-ray device. Annealed ferrite reference powders were used to quantify the instrument precision, and the accuracy of the stress measurement was tested by in situ tensile loading on 1018 carbon steel and 6061 aluminium alloy bar samples. The results show that the precision and accuracy are sensitive to the instrument (or sample) tilt angle (ψ0) as well as to the selected hkl reflection of the sample. The instrument, sample and data analysis methods all affect the overall uncertainty, and each contribution is described for this specific portable X-ray system. Finally, on the basis of the conclusions reached, desirable measurement/analysis protocols for accurate stress assessments are also presented.Keywords
This publication has 29 references indexed in Scilit:
- Stress analysis on Canadian naval platforms using a portable miniature X-ray diffractometerPowder Diffraction, 2010
- On the proper selection of reflections for the measurement of bulk residual stresses by diffraction methodsActa Materialia, 2003
- Finite element modeling of friction stir welding—thermal and thermomechanical analysisInternational Journal of Machine Tools and Manufacture, 2003
- Portable X-Ray Apparatus for Stress MeasurementsMaterials Science Forum, 2003
- Self-consistent modelling of the plastic deformation of f.c.c. polycrystals and its implications for diffraction measurements of internal stressesActa Materialia, 1998
- Imaging plate illuminates many fieldsNature, 1988
- The influence of multiaxial stress states, stress gradients and elastic anisotropy on the evaluation of (Residual) stresses by X-raysJournal of Applied Crystallography, 1979
- “PARS”—A Portable X-Ray Analyzer for Residual StressesJournal of Testing and Evaluation, 1978
- On some variational principles in anisotropic and nonhomogeneous elasticityJournal of the Mechanics and Physics of Solids, 1962
- Berechnung der elastischen Konstanten des Vielkristalls aus den Konstanten des EinkristallsThe European Physical Journal A, 1958