Tensile Strength of Cementing Agents on the CeraOne System of Dental Prosthesis on Implants
Authors:
- Alexandre Campos MontenegroAMAlexandre Campos Montenegro
- Aldir Nascimento MachadoAMAldir Nascimento Machado
- Cresus Vinicius Depes GouvêaCGCresus Vinicius Depes Gouvêa
- 1 December 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Implant Dentistry
- Vol. 17 (4) , 451-460
Abstract
The aim of this in vitro study was to evaluate the tensile strength of titanium cylinders cemented on stainless steel abutment mock-ups by the Cerazone system. Four types of cements were used: glass ionomer, Fuji I (GC); zinc phosphate, Cimento LS (Vigodent); zinc oxide without eugenol, Rely x Temp NE (3M ESPE); and resin cement, Rely x ARC (3M ESPE). Four experimental groups were formed, each composed of 5 test specimens. Each test specimen consisted of a set of 1 cylinder and 1 stainless steel abutment mock-up. All cements tested were manipulated in accordance with manufacturers' instructions. A static load of 5 Newtons (N) for 2 minutes was used to standardize the procedure. The tensile tests were performed by a mechanical universal testing machine (EMIC DL500MF) at a crosshead speed of 0.5 mm/min. The highest bonding values resulting from the experiment were obtained by Cimento LS (21.86 MPa mean), followed by the resin cement Rely x ARC (12.95 MPa mean), Fuji I (6.89 MPa mean), and Rely x Temp NE (4.71 MPa mean). The results were subjected to analysis of variance (ANOVA) and the Student's t test. The cements differed amongst them as regards tensile strength, with the highest bonding levels recorded with zinc phosphate (Cimento LS) and the lowest with the zinc oxide without eugenol (Rely x Temp NE).Keywords
This publication has 9 references indexed in Scilit:
- The comparison of provisional luting agents and abutment surface roughness on the retention of provisional implant-supported crownsThe Journal of Prosthetic Dentistry, 2006
- Effects of abutment size and luting cement type on the uniaxial retention force of implant-supported crownsThe Journal of Prosthetic Dentistry, 2000
- Tensile bond strengths of provisional luting agents used with an implant systemThe Journal of Prosthetic Dentistry, 1999
- Cement-retained versus screw-retained implant restorations: Achieving optimal occlusion and esthetics in implant dentistryThe Journal of Prosthetic Dentistry, 1997
- FACTORS INFLUENCING RETENTION OF A CeraOne GOLD CYLINDERImplant Dentistry, 1996
- THE EFFECT OF CEMENTING AGENT AND TECHNIQUE ON THE RETENTION OF A CeraOne GOLD CYLINDERImplant Dentistry, 1995
- Use of luting agents with an implant system: Part IThe Journal of Prosthetic Dentistry, 1992
- Dental CementsDental Clinics of North America, 1983
- A 15-year study of osseointegrated implants in the treatment of the edentulous jawInternational Journal of Oral Surgery, 1981