Detection of Methamphetamine in the Presence of Nicotine Using In Situ Chemical Derivatization and Ion Mobility Spectrometry
- 14 January 2004
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 76 (4) , 985-991
Abstract
The detection of methamphetamine in the presence of nicotine has been successfully accomplished using in situ chemical derivatization with propyl chloroformate as the derivatization reagent and ion mobility spectrometry (IMS). The rapid detection of methamphetamine is important for forensic scientists in order to establish a chain of evidence and link criminals to the crime scene. Nicotine is pervasive in clandestine drug laboratories from cigarette smoke residue. It has been demonstrated that nicotine obscures the methamphetamine peaks in ion mobility spectrometers due to their similar charge affinities and ion mobilities, which makes their detection a challenging task. As a consequence, false positive or negative responses may arise. In situ chemical derivatization poses as a sensitive, accurate, and reproducible alternative to remove the nicotine background when detecting nanogram amounts of methamphetamine. The derivatization agent was coated onto the sample disk, and the derivatization product corresponding to propyl methamphetamine carbamate was detected. In the present study, in situ chemical derivatization was demonstrated to be a feasible method to detect methamphetamine hydrochloride as the carbamate derivative, which was baseline-resolved from the nicotine peak. Alternating least squares (ALS) was used to model the datasets. A mixture containing both compounds revealed reduced mobilities of 1.61 cm2/V·s and 1.54 cm2/V·s for methamphetamine and nicotine, respectively. The reduced mobility of propyl methamphetamine carbamate was found at 1.35 cm2/V·s.Keywords
This publication has 22 references indexed in Scilit:
- Analysis of methamphetamine and its metabolites in hairBiomedical Chromatography, 2003
- Analysis of 2,4,6-trinitrotoluene, pentaerythritol tetranitrate and cyclo-1,3,5-trimethylene-2,4,6-trinitramine using negative corona discharge ion mobility spectrometryTalanta, 2003
- Application of equality constraints on variables during alternating least squares proceduresJournal of Chemometrics, 2002
- Peer Reviewed: Teaching Analytical Chemistry through Mineral AnalysisAnalytical Chemistry, 2002
- Field detection and monitoring of explosivesTrAC Trends in Analytical Chemistry, 2002
- Determination of soman and VX degradation products by an aspiration ion mobility spectrometryAnalytica Chimica Acta, 2001
- Ion Mobility Spectrometry: Arriving on Site and Moving beyond a Low ProfileApplied Spectroscopy, 1999
- Comparison between the direct trilinear decomposition and the multivariate curve resolution-alternating least squares methods for the resolution of three-way data setsChemometrics and Intelligent Laboratory Systems, 1998
- Ion mobility spectrometric studies of organophosphorus compoundsAnalytica Chimica Acta, 1992
- Interactive self-modeling mixture analysisAnalytical Chemistry, 1991