Abstract
OBJECTIVE. The purpose of this article is to assess detection, tracking, and reading time of solid lung nodules ≥ 4 mm on pairs of MDCT chest screening examinations using a computer-aided detection (CAD) system.MATERIALS AND METHODS. Of 54 pairs of low-dose MDCT chest examinations (1.25-mm collimation), two chest radiologists in consensus established that 25 examinations contained 52 nodules ≥ 4 mm. All paired examinations were interpreted on the CAD workstation—first without and then with CAD input—for the detection and tracking of lung nodules. A subset of 33 examination pairs was later read on the clinical workstation used in daily practice, and the results were compared for reading time with those on the CAD workstation.RESULTS. After CAD input, the sensitivity for nodule detection increased statistically significantly for both readers (9.6% and 23%; p ≤ 0.025). One cancer initially missed by one radiologist was correctly identified with CAD input. The overall reading time on the CAD workstation and c...