Long‐Term Outcome of Leads and Patients Following Robotic Epicardial Left Ventricular Lead Placement for Cardiac Resynchronization Therapy
Authors:
- 28 October 2010
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 34 (2) , 235-240
Abstract
Introduction: In cardiac resynchronization therapy (CRT), positive clinical response and reverse remodeling have been reported using robotically assisted left ventricular (LV) epicardial lead placement. However, the long‐term performance of epicardial leads and long‐term outcome of patients who undergo CRT via robotic assistance are unknown. In addition, since the LV lead placement is more invasive than a transvenous procedure, it is important to identify patients at higher risk of complications. Methods: We evaluated 78 consecutive patients (70 ± 11 years, 50 male) who underwent robotic epicardial LV lead placement. The short‐ ( Results: All patients had successful lead placement and were discharged in stable condition. When compared to the time of implantation, there was a significant increase in pacing threshold (1.0 ± 0.5 vs 2.14 ± 1.2; P < 0.001) and decrease in lead impedance (1010 ± 240 Ω vs 491 ± 209 Ω; P < 0.001) at short‐term follow‐up. The pacing threshold (2.3 ± 1.2 vs 2.14 ± 1.2; P = 0.30) and lead impedance (451 ± 157 Ω vs 491 ± 209 Ω; P = 0.10) remained stable during long‐term follow‐up when compared to short‐term values. At a follow‐up of 44 ± 21 months, there were 20 deaths (26%). These patients were older (77 ± 7 vs 67 ± 11 years; P = 0.001) and had a lower ejection fraction (EF) (13 ± 7% vs 18 ± 9%; P = 0.02) than surviving patients. Conclusion: Robotically implanted epicardial LV leads for CRT perform well over short‐ and long‐term follow‐up. Older patients with a very low EF are at higher risk of death. The risks and benefits of this procedure should be carefully considered in these patients. (PACE 2011; 34:235–240)Keywords
This publication has 13 references indexed in Scilit:
- A prospective longitudinal evaluation of the benefits of epicardial lead placement for cardiac resynchronization therapyEP Europace, 2009
- Mortality of Heart Failure Patients After Cardiac Resynchronization Therapy: Identification of PredictorsJournal of Cardiovascular Electrophysiology, 2008
- Very Long Term Follow-Up of Cardiac Resynchronization Therapy: Clinical Outcome and Predictors of MortalityEuropean Journal of Heart Failure, 2008
- Epicardial versus Transvenous Left Ventricular Lead Placement in Patients Receiving Cardiac Resynchronization Therapy: Results from a Randomized Prospective StudyThe Thoracic and Cardiovascular Surgeon, 2008
- Follow-Up of Robotically Assisted Left Ventricular Epicardial Leads for Cardiac Resynchronization TherapyJACC, 2005
- The Effect of Cardiac Resynchronization on Morbidity and Mortality in Heart FailureNew England Journal of Medicine, 2005
- Surgical epicardial left ventricular lead versus coronary sinus lead placement in biventricular pacing☆European Journal of Cardio-Thoracic Surgery, 2005
- Comparison of response to cardiac resynchronization therapy in patients with sinus rhythm versus chronic atrial fibrillationThe American Journal of Cardiology, 2004
- Comparison of left ventricular lead placement via the coronary venous approach versus lateral thoracotomy in patients receiving cardiac resynchronization therapyThe American Journal of Cardiology, 2004
- Robotically assisted left ventricular epicardial lead implantation for biventricular pacingJACC, 2003