In Vitro Testing of a Novel Blood Pump Designed for Temporary Extracorporeal Support
- 1 March 2012
- journal article
- adult circulatory-support
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 58 (2) , 109-114
Abstract
Extracorporeal blood pumps are used as temporary ventricular assist devices or for extracorporeal membrane oxygenation. The ideal pump would be intrinsically self-regulating, carry no risk of cavitation or excessive inlet suction, be afterload insensitive, and valveless thus reducing thrombogenicity. Currently used technology, including roller, centrifugal, and pneumatic pulsatile pumps, does not meet these requirements. We studied a nonocclusive peristaltic pump (M-Pump) in two mock circulatory loops and compared the performance to a frequently used centrifugal pump and a modified prototype of the M-Pump (the BioVAD). The simple resistance loop consisted of the investigated pump, a fixed height reservoir at 150 mm Hg, and a variable inflow reservoir. The pulsatile circulation used a mock patient simulator with adjustable resistance elements connected to a pneumatic pulsatile pump. The M-Pump intrinsically regulated flow with changing preload, was afterload insensitive, and did not cavitate, unlike the centrifugal pump. The BioVAD also demonstrated these features and could augment output with the use of vacuum assistance. A nonocclusive peristaltic pump may be superior for short-term extracorporeal circulatory assist by mitigating risks of excessive inlet suction, afterload sensitivity, and thrombosis.Keywords
This publication has 10 references indexed in Scilit:
- Results of the Post-U.S. Food and Drug Administration-Approval Study With a Continuous Flow Left Ventricular Assist Device as a Bridge to Heart TransplantationJACC, 2011
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceNew England Journal of Medicine, 2009
- Short- and Long-Term Survival of Patients Transferred to a Tertiary Care Center on Temporary Extracorporeal Circulatory SupportThe Annals of Thoracic Surgery, 2009
- Extracorporeal Membrane Oxygenation to Support Cardiopulmonary Resuscitation in AdultsThe Annals of Thoracic Surgery, 2009
- Impact of Center Volume on Outcomes of Left Ventricular Assist Device Implantation as Destination TherapyCirculation: Heart Failure, 2009
- Cardiogenic ShockCirculation, 2008
- “Bridge to Decision”—Time to Consider an Additional Management ParadigmThe Journal of Heart and Lung Transplantation, 2008
- Extracorporeal Life Support for Massive Pulmonary EmbolismJournal of Trauma: Injury, Infection & Critical Care, 2007
- Significant Safety Advantages Gained with an Improved PressureRegulated Blood PumpThe Journal of ExtraCorporeal Technology, 1996
- Laboratory Experience with a Novel, Non-occlusive, Pressure-regulated Peristaltic Blood PumpASAIO Journal, 1992